Owner's Operator and
Maintenance Manual
DEMAND OXYGEN DELIVERY
DEVICE AND SYSTEM
Venture 20EX, 20 PSIG Device
Venture 50EX, 50 PSIG Device
Venture Pack 211EX, 20 PSIG Shoulder Bag System
w/M6 Cylinder
Venture Pack 212EX, 20 PSIG Shoulder Bag System
w/M9 Cylinder
DEALER: THIS MANUAL MUST BE GIVEN TO THE USER OF THIS
DEMAND OXYGEN DELIVERY DEVICE.
USER: BEFORE USING THIS DEMAND OXYGEN DELIVERY
DEVICE, READ THIS MANUAL AND SAVE FOR FUTURE REFERENCE.
WARNING
CAUTION
???Caution: Federal law restricts this device to sale by or on order of a physician, or any other practitioner licensed by the law of the State in which he practices to use or order the use of this device.???
WARNING
DO NOT install, maintain or operate this equipment without first reading and understanding this manual. If you are unable to understand the WARNINGS and INSTRUCTIONS, contact your Home Health Care Dealer before attempting to use this equipment - otherwise injury or damage may result. This includes instructions that pertain to the:
Oxygen Cylinder,
Oxygen Regulator,
Venture Demand Oxygen Delivery Device, and/or Liquid Oxygen System
WARNING/CAUTION notices as used in this manual apply to hazards or unsafe practices which could result in personal injury or property damage.
NOTICE
The information contained in this document is subject to change without notice.
SAVE THESE INSTRUCTIONS
SAFETY SUMMARY
WARNING
This device should ONLY be used under the supervision of a physician. Verify your oxygen flowrate with your physician when you begin using your Demand Oxygen Delivery System.
If you have any questions about the oxygen delivery system or its use, call your Home Health Care Dealer or Service Representative.
If you suspect that an oxygen delivery system has been damaged, contact your Service Representative IMMEDIATELY. Your Service Representative MUST handle all maintenance.
Keep all oxygen delivery systems out of the reach of children.
Avoid creation of any spark near medical oxygen equipment. This includes sparks from static electricity created by any type of friction.
All electrical equipment near oxygen units MUST be properly grounded.
This system only delivers medical grade (U.S.P.) oxygen when used with oxygen cylinders marked as such. When using this system with Invacare Venture Homefill, oxygen delivery will be USP 93%+3%. Oxygen supplied by the system is supplemental and is NOT to be considered
Before using the Demand Oxygen Delivery Device, check the contents gauge of the oxygen supply
equipment. The oxygen supply equipment MUST be filled only by an authorized Service Representative, unless being filled by the Invacare Venture Homefill system.
If at anytime the attendant or the end user concludes that an insufficient amount of oxygen is being received,
SAFETY SUMMARY
WARNING (CONTINUED)
the supplier and/or physician should be contacted IMMEDIATELY. No adjustments should be made to the flowrate unless prescribed by a physician.
DO NOT attempt to alter the switches in the battery compartment. These switches are preset by your Home Health Care Dealer and control the breath delivery prescription.
RISK OF FIRE:
Oxygen does not burn. However, it will accelerate burning of flammable materials.
TO PREVENT FIRE:
DO NOT SMOKE when operating the oxygen delivery system. No Smoking signs should be prominently displayed.
Keep oxygen delivery system at least
Keep the oxygen delivery system away from open flames or heating sources such as furnaces or stoves.
Keep flammable materials away from the oxygen delivery system; such as oils, grease and aerosol sprays. These will ignite and burn quickly in the presence of concentrated oxygen.
TO PREVENT HIGH CONCENTRATIONS OF OXYGEN: Keep the oxygen delivery system in a well ventilated area at all times.
When the oxygen delivery system is not in use, press the OFF button to place the Demand Oxygen Delivery Device in the Continuous Mode, turn the Cylinder OFF, bleed the remaining oxygen from the Regulator, turn the Regulator OFF.
SAFETY SUMMARY
WARNING (CONTINUED)
DO NOT wear the Demand Oxygen Delivery Device under any clothing.
OTHER PRECAUTIONS:
Never attempt to lubricate the oxygen delivery system.
DO NOT insert any object (except suitable AC/DC Adapter) in external power jack of Demand Oxygen Delivery Device. This may cause an electrical shock which could cause burns.
DO NOT subject the Demand Oxygen Delivery Device to temperatures below
NEVER tamper with the oxygen delivery system. Doing so could create a hazardous condition and
possibly cause the oxygen delivery system to malfunction. If you have any questions, call your Home Health Care Dealer or Service Representative.
ALWAYS connect the Demand Oxygen Delivery Device to a pressure regulated oxygen source. ONLY connect 20 psig Demand Oxygen Delivery Device to a 20 psig oxygen source and a 50 psig Demand Oxygen Delivery Device to a 50 psig oxygen source.
DO NOT use any additional lengths of tubing between the oxygen source and the Demand Oxygen Delivery Device. ONLY use Invacare tubing supplied with the Demand Oxygen Delivery Device, if any tubing is required.
DO NOT use a mask when using pulse mode delivery,
USE ONLY a standard nasal cannula.
When using the continuous flow delivery, a mask or nasal cannula can be used.
SAFETY SUMMARY
WARNING (CONTINUED)
DO NOT use a humidifier with the Demand Oxygen Delivery Device.
DO NOT expose the Demand Oxygen Delivery Device or system to water or any other liquids.
RADIO FREQUENCY INTERFERENCE
Most electronic equipment is influenced by Radio Frequency Interference (RFI). CAUTION should be exercised with regard to the use of portable communications equipment in the area around such equipment.
UNPACKING / HANDLING
UNPACKING
1.Check for any obvious damage to the carton or its contents. (If damage is evident, notify your local Invacare Home Health Care Dealer/Carrier.)
2.Remove all the loose packing from the carton.
3.Carefully remove all the components from the carton.
The following is a listing of items included in the device and system packages:
HANDLING PROCEDURES
NOTE: Unless the Venture Demand Oxygen Delivery Device is to be used immediately, retain boxes, containers and packing materials for use in returning or storing the equipment until usage is required.
INSPECTION
1.Examine exterior of the equipment for nicks, dents, scratches or other damages. Inspect all components.
STORAGE
1.Carefully repackage all components in the original packaging.
2.Store the repackaged equipment in a dry area.
3.DO NOT place other objects on top of the repackaged equipment.
INTRODUCTION
INVACARE VENTURE DEMAND OXYGEN
DELIVERY DEVICE
Ordinary oxygen supply units deliver oxygen continuously.
The normal breathing pattern is to inhale about
In contrast, the Venture Demand Oxygen Delivery Device (Venture DODD) senses your inhalation and instantly releases a programmed pulse of oxygen at a specified flowrate at the beginning of inhalation.
The specified flowrate of oxygen is delivered within the first second of inhalation. Oxygen is not delivered during exhalation. This means that your oxygen use time is extended by the ratio of breath duration over the pulse duration. Venture DODD will extend user time an average of four (4) times in the Variable duration breath mode.
Comfort is another advantage of the Venture DODD. Nasal drying caused by continuous flow is eliminated, as oxygen is only delivered during inhalation. Therefore, additional humidification is not necessary. Since the oxygen is delivered at normal flowrates, there is not a high blast of oxygen to the end user, which makes the Venture DODD more comfortable than other pulse systems.
The Venture DODD can be used in the traditional continuous flow mode simply by turning the Venture DODD device OFF. Effective flow deliveries of 0.5 to 6 liters per minute (L/min.) can be instantly selected, in pulse or continuous mode, by each individual as prescribed by your physician.
During pulse mode, the Venture DODD runs on a disposable
(Refer to the OPERATION section in this manual for information on the Low Battery Indicator LED). During continuous mode operation, the unit does not need the battery to operate.
NOTE: Actual Battery use time depends on your breathing rate.
The Venture DODD can be used with compressed gas or liquid oxygen systems. Both 20 and 50 psig units were designed for use on 20 and 50 psig oxygen delivery systems, respectively.
NOTE: The 50 psig system should ONLY be used with the threaded (DISS) fitting, and NOT the
FEATURES
VENTURE DEMAND OXYGEN DELIVERY
DEVICE COMPONENTS AND DEFINITION
OF TERMS (FIGURE 1)
ON/OFF Buttons - Determines flow selection method. The ON button turns Power ON for normal pulse operation ( ) and the OFF button disengages the electronics for continuous flow operation ( ).
NOTE: The OFF button does NOT stop flow delivery and is NOT recommended for unit shutdown. See SETUP AND NORMAL PULSE OPERATION in the OPERATION Section of this manual.
Oxygen OUTLET Fitting - The delivery point of oxygen to the end user from the Venture DODD. The fitting is barbed to securely hold the cannula for oxygen delivery.
Oxygen INLET Fitting - The delivery point of oxygen from the oxygen system to the Venture DODD.
External Power Input Jack - AC/DC Power Adapter connection port.
Status Indicators:
O2 - GREEN LED will illuminate with each breath indicating that the Venture DODD is operating.
- Attention - YELLOW LED illumination indicates the Venture DODD is NOT sensing end user's inhalation or that there is an internal malfunction.
Refer to TROUBLESHOOTING Section.
- Low Battery - RED LED indicates two (2) low battery
levels.
O2 - Low/No Gas - GREEN LED illumination indicates Venture DODD is NOT sensing O2 presence in pulse delivery.
FEATURES
??? External Power
Input Jack
??? ???
Oxygen OUTLET Fitting Oxygen INLET Fitting
Low O2 LED (GREEN)
Battery LED (RED)
Attention LED (YELLOW)
Pulse LED (GREEN)
ON(Pulse ModeOperation)
OFF (Continuous Mode Operation)
NOTE: Battery Compartment located on bottom back of Venture Demand Oxygen ??? Delivery Device.
FIGURE 1 - VENTURE DODD FEATURES
FEATURES
VENTURE DODD SYSTEM COMPONENTS AND
DEFINITION OF TERMS (FIGURE 2)
Cylinder - Lightweight aluminum cylinder provided.
Regulator - Provides flowrate control of oxygen delivery with compressed gas cylinders.
DISS Fitting - Connection point for direct connect to oxygen source.
Barb Fitting - Connection point for tubing used with compressed gas cylinders.
Connection Tubing - Provides oxygen path from regulator to the device inlet barb fitting for use with compressed gas cylinders.
Carrying Case with Padded Strap - Used for convenient transportation and is adjustable to ensure comfort.
Optional AC/DC Adpater - Designed to provide AC current to device.
FIGURE 2 - VENTURE DODD SYSTEM FEATURES
SPECIFICATIONS
OPERATION
THE USE OF OXYGEN
The type of oxygen administrating equipment to be used, as well as the flowrate and time of usage, must be prescribed by a physician. Your physician should verify your Venture DODD flowrate prescription.
WARNING
If at anytime the attendant or the end user concludes that an insufficient amount of oxygen is being received, the supplier and/or physician should be contacted im- mediately. No adjustments should be made to the flowrate unless prescribed by a physician.
NOTE: ONLY a NASAL CANNULA can be used in the PULSE MODE to deliver oxygen. An oxygen mask or nasal cannula can be used with Continuous Flow to deliver oxygen. Low flow cannula should not be used.
1.Install the battery into the Venture DODD or plug AC/DC power adapter into external power jack.
A.To Install or Replace the Battery:
Open the battery cover on the back of the Venture DODD by applying light inward pressure with your thumb to the cover latch and then gently lift the cover open.
If replacing a battery: Remove the used battery from the battery compartment by grasping the top and bottom ends and pulling straight out.
To install the battery: Orient the new battery so that its polarity matches the polarity marked in the battery compartment.
WARNING
DONOTattempttoaltertheswitchesinthebattery compartment.Theseswitchesarepresetbyyourhome healthcareproviderandcontrolthebreathdeliveryprescription.
Press the new battery in place and close the battery cover.
2. Check Oxygen Source to verify adequate oxygen supply.
OPERATION
3.If using an oxygen cylinder, check that the regulator is properly attached before opening the cylinder valve.
4.Connect the Venture DODD to the regulator by performing one (1) of the following:
A.If using an Oxygen Cylinder and Regulator:
On a 20 psig regulator, connect the tubing (supplied with the Venture DODD) from the barb fitting of the regulator to the barb inlet fitting on the Venture DODD.
On a 50 psig regulator, connect the threaded (DISS) fitting (supplied with the Venture DODD) to the regulator DISS fitting.
NOTE: The 50 psig system should ONLY be used with the threaded (DISS) fitting and not the
B.If using a Liquid Oxygen Reservoir:
Directly connect the Venture DODD to the oxygen source using a DISS fitting.
5.Connect the nasal cannula tubing to the barbed oxygen outlet fitting on the Venture DODD and adjust cannula for comfort.
WARNING
When using nasal or masked type administering equipment, oxygen or air tubing MUST be routed and secured properly to ensure that the tubing does NOT become entangled, kinked or severed.
NOTE: Use ONLY a NASAL CANNULA with PULSE MODE
delivery, NOT a mask. A mask may NOT fit your face well enough for the unit to sense your inhalation.
NOTE: A special nasal cannula is NOT required. Any standard flow
NOTE: A MASK or a CANNULA may be used with CONTINUOUS FLOW.
NOTE: The cannula tubing can be up to 50 feet long to function properly in the pulse mode. The continuous flow mode is NOT affected by the length of the canula tubing.
OPERATION
6.Set the prescribed flow rate on the oxygen source by performing one (1) of the following:
A.If using an Oxygen Cylinder and Regulator:
OPEN cylinder valve and adjust regulator to prescribed flow rate.
B.If using a Liquid Oxygen Reservoir:
Set flow control knob on reservoir to prescribed flowrate.
7.Press the ON button to engage pulse mode. The unit will pulse each time the user inhales, the GREEN PULSE LED will blink with each oxygen pulse delivery.
8.As battery power depletes, the unit will alarm to inform the end user to replace the battery or switch to external power source.
A.INITIAL BATT. LED - The Red LED will FLASH and the audible alarm will BEEP three (3) times. This indicates that you have about two (2) hours of use time left on your battery.
B.SECOND BATT. LED - The Red LED will be SOLID and the audible alarm will BEEP six (6) times. This indicates that you have between 15 and 30 minutes of use time left on your battery.
NOTE: Remaining battery use times are dependent on battery type being used. Alkaline battery type is recommended for longest life.
OFF
(Continuous Mode
Operation)
FIGURE 3 - SWITCHING TO CONTINUOUS FLOW
OPERATION
NOTE: Whenever the pulse mode is used, the battery is monitored. If the SECOND BATT. LED illuminates, either replace the battery or switch to continuous mode if oxygen supply is adequate. When in continuous mode, the battery is NOT used.
NOTE: When in the continuous flow mode, the oxygen supply will be used considerably faster then in the pulse mode. If using a portable oxygen supply, make plans for an alternative source of oxygen. Replace the battery as soon as possible.
9.If it is necessary to switch to continuous flow at any time, press the OFF button (FIGURE 3).
10.If turning the system OFF for cylinder replacement or storage, press the Venture DODD OFF button. Turn OFF the cylinder post valve, bleed the oxygen from the regulator and turn OFF the regulator prior to any disassembly.
MAINTENANCE
PROLONGING THE LIFE OF YOUR BATTERY
The Venture DODD battery is powered by a single
NOTE: The Alkaline battery should last for approximately 80 hours with normal use, but will last even longer if you follow this procedure:
1.Use the battery to its fullest capacity prior to replacing. This is accomplished by running the unit from the battery until it stops working (approximately 15 to 30 minutes after the SECOND BATT. indicator illuminates SOLID and the audible alarm sounds six [6] times).
CAUTION
DO NOT store Venture DODD with the battery installed.
WARNING
DO NOT recharge, dispose of in fire, mix with different battery type, connect improperly or disassemble Alkaline batteries. Doing so may cause battery to explode or leak resulting in injury.
2.Operate your unit from the AC/DC power adapter when at home.
3.Avoid exposure to extreme temperatures.
WARNING
DO NOT operate the unit in temperatures above 104??F (40??C) or below 40??F (4??C), or place near heat sources such as radiators or fires.
Electrical shock hazard. DO NOT remove outer case of this unit. All disassembly and maintenance of this unit MUST BE done by an Invacare Dealer. If outer case is removed by anyone EXCEPT an Invacare Dealer the warranty will be voided.
MAINTENANCE
CARE OF YOUR VENTURE DEMAND
OXYGEN DELIVERY DEVICE AND SYSTEM
CLEANING THE OUTER CASE
WARNING
DO NOT submerge in water to clean.
DO NOT open outer case. If outer case is removed by anyone EXCEPT an Invacare Dealer the warranty will be voided.
CAUTION
Use no oil. Never apply waxes, polishes, or other similar compounds to any surface of the Venture Demand Oxygen Delivery Device, as these compounds may be flammable.
Clean outer case with damp cloth. If necessary, a mild dish detergent may be used. Dry outer case with lint free cloth.
CARRYING CASE
The carrying case can be hand washed in warm water. Air dry ONLY. NO bleach. DO NOT dryclean.
ACCESSORIES AND SPARE PARTS
M6 Aluminum Cylinder, 164L Capacity
M9 Aluminum Cylinder, 248L Capacity
Piston Regulator,
(Dual size for Cylinder diameters of 3 and
Venture Operator's Manual
NOTE: Rechargeable batteries can be used to power Venture DODD, however the batteries and charger are a user supply item not provided by Invacare.
NOTE: If you have any questions regarding accessories, please call Invacare or your Home Healthcare Dealer.
TROUBLESHOOTING
TROUBLESHOOTING
TROUBLESHOOTING
NOTE: If any of the above problems occur and the solutions do not work, please contact your Invacare Home Health Care Dealer for service.
OXYGEN CONSERVATION
OXYGEN CONSERVED AND BATTERY
USE TIME BASED ON BREATH RATE
i.e. 77% Oxygen conservation over Continuous Flow achieved with Venture DODD in Variable Breath Mode @ 20 BPM Rate. Oxygen supply lasts
COMPATIBILITY
Venture DODD requires using a flowrate controlled oxygen source. Gaseous or liquid oxygen systems that meet the following requirements are compatible to the Venture DODD:
Venture 20 requires a 20 +3 psig (138 + 20 kPa) flow source. Venture 50 requires a 50 +5 psig (345 + 35 kPa) flow source. Flowrate Range:
LIMITED WARRANTY
PLEASE NOTE: THE WARRANTY BELOW HAS BEEN DRAFTED
TO COMPLY WITH FEDERAL LAW APPLICABLE TO PRODUCTS MANUFACTURED AFTER JULY 4, 1975.
This warranty is extended only to the original purchaser/user of our products.
This warranty gives you specific legal rights and you may also have other legal rights which vary from state to state.
Invacare warrants the Demand Oxygen Delivery Device to be free from defects in materials and workmanship for a period of one (1) year from date of purchase. In addition, all internal electronic components and switches carry a three (3) year warranty from date of purchase. If within such warranty period any such product shall be proven to Invacare satisfaction to be defective, product shall be repaired or replaced, at Invacare option. This warranty does not include normal wear and tear or shipping charges incurred in replacement part installation or repair of such product. Invacare sole obligation and your exclusive remedy under this warranty shall be limited to such repair and/or replacement.
For warranty service, please contact Invacare Service Department at the Toll Free Number listed below during normal hours.
Upon receiving notice of an alleged defect in a product, Invacare will issue a serialized return authorization. It shall be the responsibility of the purchaser to return, at the purchaser???s cost, to either Invacare???s plant or service center as specified by Invacare in advance. Defective components part(s) must be returned for warranty inspection using the serial number for identification within 30 days of the return authorization date. DO NOT return products to our factory without prior consent. C.O.D. shipments will be refused - please prepay shipping charges.
LIMITATIONS AND EXCLUSIONS: THE FOREGOING
WARRANTY SHALL NOT APPLY TO SERIAL NUMBERED PROD-
UCTS IF THE SERIAL NUMBER HAS BEEN REMOVED OR DE-
FACED, PRODUCTS SUBJECTED TO NEGLIGENCE,
ACCIDENT, IMPROPER OPERATION, MAINTENANCE OR STOR-
AGE, PRODUCTS MODIFIED WITHOUT INVACARE???S EXPRESS
WRITTEN CONSENT (INCLUDING, BUT NOT
LIMITED TO, MODIFICATION THROUGH THE USE OF
UNAUTHORIZED PARTS OR ATTACHMENTS; PRODUCTS
DAMAGED BY REASON OF REPAIRS MADE TO ANY
COMPONENT WITHOUT THE SPECIFIC CONSENT OF
INVACARE, OR TO A PRODUCT DAMAGED BY
CIRCUMSTANCES BEYOND INVACARE RESPIRATORY -
LIMITED WARRANTY
CORPORATION???S CONTROL, AND SUCH EVALUATION WILL BE
SOLELY DETERMINED BY INVACARE. THE WARRANTY SHALL
NOT APPLY TO PROBLEMS ARISING FROM NORMAL WEAR OR
FAILURE TO ADHERE TO THE FOLLOWING INSTRUCTIONS.
THE FOREGOING EXPRESS WARRANTY IS EXCLUSIVE AND IN
LIEU OF ANY OTHER WARRANTIES WHATSOEVER,
WHETHER EXPRESS OR IMPLIED, INCLUDING THE IMPLIED
WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A
PARTICULAR PURPOSE, AND THE SOLE REMEDY FOR
VIOLATIONS OF ANY WARRANTY WHATSOEVER, SHALL BE
LIMITED TO REPAIR OR REPLACEMENT OF THE DEFECTIVE
PRODUCT PURSUANT TO THE TERMS CONTAINED HEREIN.
THE APPLICATION OF ANY IMPLIED WARRANTY
WHATSOEVER SHALL NOT EXTEND BEYOND THE DURATION
OF THE EXPRESS WARRANTY PROVIDED HEREIN.
INVACARE SHALL NOT BE LIABLE FOR ANY CONSEQUENTIAL
OR INCIDENTAL DAMAGES WHATSOEVER.
SOME STATES DO NOT ALLOW THE EXCLUSION OR
LIMITATION OF INCIDENTAL OR CONSEQUENTIAL DAMAGE,
OR LIMITATION ON HOW LONG AN IMPLIED WARRANTY
LASTS, SO THE ABOVE EXCLUSION AND LIMITATION MAY
NOT APPLY TO YOU.
THIS WARRANTY SHALL BE EXTENDED TO COMPLY WITH
STATE/PROVINCIAL LAWS AND REQUIREMENTS.
_____________________________________________________________________________________________________________________