Performance Tests and
Adjustments Manual
HP 8568B Spectrum Analyzer
HEWLETT
PACKARD
HP Part No.
Printed in USA September 1993
@Copyright
All Rights Reserved. Reproduction, adaptation, or translation without prior written permission is prohibited, except as allowed under the copyright laws.
1212 Valley House Drive, Rohnert Park, CA
Certification
Warranty
This
For warranty service or repair, this product must be returned to a service facility designated by
its programming instructions when properly installed on that instrument.
LIMITATION OF WARRANTY
The foregoing warranty shall not apply to defects resulting from improper or inadequate maintenance by Buyer,
NO OTHER WARRANTY IS EXPRESSED OR IMPLIED.
WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A
PARTICULAR PURPOSE.
EXCLUSIVE REMEDIES
THE REMEDIES PROVIDED HEREIN ARE BUYER???S SOLE AND
EXCLUSIVE REMEDIES.
LIABLE FOR ANY DIRECT, INDIRECT, SPECIAL, INCIDENTAL, OR
CONSEQUENTIAL DAMAGES, WHETHER BASED ON CONTRACT,
TORT, OR ANY OTHER LEGAL THEORY.
. . .
III
Assistance
Product maintenance agreements and other customer assistance agreements are available for
Fbr any assistance, contact your nearest
Safety Symbols
The following safety symbols are used throughout this manual. Familiarize yourself with each of the symbols and its meaning before operating this instrument.
Caution The caution sign denotes a hazard. It calls attention to a procedure which, if not correctly performed or adhered to, could result in damage to or destruction of the instrument. Do not proceed beyond a caution sign until the indicated conditions are fully understood and met.
Warning The warning sign denotes a hazard. It calls attention to a procedure which, if not correctly performed or adhered to, could result in injury or loss of life. Do not proceed beyond a warning sign until the indicated conditions are fully understood and met.
General Safety
Considerations
Warning Before this instrument is switched on, make sure it has been properly grounded through the protective conductor of the ac power cable to a socket outlet provided with protective earth contact.
Any interruption of the protective (grounding) conductor, inside or outside the instrument, or disconnection of the protective earth terminal can result in personal injury.
Warning There are many points in the instrument which can, if contacted, cause personal injury. Be extremely careful.
Any adjustments or service procedures that require operation of the instrument with protective covers removed should be performed only by trained service personnel.
Caution Before this instrument is switched on, make sure its primary power circuitry has been adapted to the voltage of the ac power source.
Failure to set the ac power input to the correct voltage could cause damage to the instrument when the ac power cable is plugged in.
V
HP 8568B
Spectrum Analyzer
Documentation
Outline
HP 8568B Installation
and Verification
Manual
HP 8568B Operating
and Programming
Manual
HP 8568B
Performance Tests and
Adjustments Manual
Included with the HP Model 8568B Spectrum Analyzer are three manuals: the Installation and Verification Manual, the Operating and Programming Manual, and the Performance Tests and Adjustments Manual.
General information, installation, specifications, characteristics, and operation verification.
Manual and remote operation, including complete syntax and command description. Accompanying this manual is the separate,
Electrical performance tests and adjustment procedures.
HP 85680B RF Section RF Section service information.
Troubleshooting and
Repair Manual
HP 85662A
Section
Troubleshooting and
Repair Manual
vi
Contents
4.Resolution Bandwidth Accuracy Test . . . . . .
5.Resolution Bandwidth Selectivity Test . . . . . .
6.Resolution Bandwidth Switching Uncertainty Test
7.Input Attenuator Switching Uncertainty Test . .
8.Frequency Response Test . . . . . . . . . . . .
9.RF Gain Uncertainty Test . . . . . . . . . . . .
10.IF Gain Uncertainty Test . . . . . . . . . . .
11.Log Scale Switching Uncertainty Test . . . . .
12.Amplitude Fidelity Test . . . . . . . . . . . .
13.Average Noise Level Test . . . . . . . . . . .
14.Residual Responses Test . . . . . . . . . . . .
15.Spurious Responses Test . . . . . . . . . . . .
16.Residual FM Test . . . . . . . . . . . . . . .
17.
18.Calibrator Amplitude Accuracy Test . . . . . .
19.Fast Sweep Time Accuracy Test (~20 ms) . . . .
20.1st LO Output Amplitude Test . . . . . . . . .
21.Frequency Reference Error Test . . . . . . . .
Test 3. Sweep Time Accuracy . . . . . . . . . . . .
Test . . . . . . . . . . . . . . . . . . . . . .
C o n t e n t s - l
Test 11. Log Scale Switching Uncertainty Test . . . .
3.Adjustments
1.
2.
2.
3.Preliminary Display Adjustments (SN 3001A and
4.Final Display Adjustments (SN 3001A and Below) .
4.Final Display Adjustments (SN 3004A and Above) .
5.Log Amplifier Adjustments . . . . . . . . . . . .
6.Video Processor Adjustments . . . . . . . . . . .
7.3 MHz Bandwidth Filter Adjustments . . . . . . .
8.21.4 MHz Bandwidth Filter Adjustments . . . . .
9.3 dB Bandwidth Adjustments . . . . . . . . . .
10.Step Gain and 18.4 MHz Local Oscillator
Adjustments . . . . . . . . . . . . . . . . . .
11.Down/Up Converter Adjustments . . . . . . . .
12.Time Base Adjustment (SN 2840A and Below, also
32 17AO5568 and Above) . . . . . . . . . . . . .
12.Time Base Adjustment (SN 2848A to 3217A05567)
13.20 MHz Reference Adjustments . . . . . . . . .
14.249 MHz Phase Lock Oscillator Adjustments . . .
15.275 MHz Phase Lock Oscillator Adjustment . . . .
16.Second IF Amplifier and Third Converter
Adjustment . . . . . . . . . . . . . . . . . . .
17.Pilot Second IF Amplifier Adjustments . . . . . .
18.Frequency Control Adjustments . . . . . . . . .
19.Second Converter Adjustments . . . . . . . . .
20.50 MHz
21. Slope Compensation Adjustments . . . . . . . .
22.Comb Generator Adjustments . . . . . . . . . .
23.
24.Track and Hold Adjustments . . . . . . . . . . .
25.Digital Storage Display Adjustments . . . . . . .
Crystal Filter Bypass Network Configuration . . . . .
4. Option 462
Introduction . . . . . . . . . . . . . . . . . . . . .
4. 6 dB Resolution Bandwidth Accuracy Test . . . . .
4.Impulse and Resolution Bandwidth Accuracy Test .
5.6 dB Resolution Bandwidth Selectivity Test . . . .
5.Impulse and Resolution Bandwidth Selectivity Test .
6.Impulse and Resolution Bandwidth Switching Uncertainty Test . . . . . . . . . . . . . . . .
Test 4. 6 dB Resolution Bandwidth Accuracy Test (p/o Table
Test 4. Impulse and Resolution Bandwidth Accuracy Test (p/o Table
Test 5. 6 dB Resolution Bandwidth Selectivity (p/o ???Ihble
Test 5. Impulse and Resolution Bandwidth Selectivity (p/o Table
Test 6. Impulse and Resolution Bandwidth Switching Uncertainty (p/o ???Iable
9. 6 dB Resolution Bandwidth Adjustments . . . . .
9.Impulse Bandwidth Adjustments . . . . . . . . .
5.Option 857
Introduction . . . . . . . . . . . . . . . . . . . . .
12. Option 857 Amplitude Fidelity Test . . . . . . .
Performance Test Record . . . . . . . . . . . . . . .
Test 12. Option 857 Amplitude Fidelity Test . . . . .
6.Major Assembly and Component Locations
RF Section Figure Index . . . . . . . . . . . . . . .
Figures
lhbles
2- 17. Linear Amplitude Fidelity . . . . . . . . . . . . . .
3.5. Initial Adjustment Positions . . . . . . . . . . . . .
1
General Information
Introduction
Warning
Instruments Covered
by this Manual
This HP 8568B Tests and Adjustments Manual contains two sections: Performance Tests and Adjustments Procedures. The Performance Tests provided should be performed for the following reasons:
w If the test equipment for the Operation Verification Program is not available.
n If the instrument does not pass all of the Operation Verification tests.
w For complete verification of specifications not covered by the Operation Verification program.
The adjustment procedures should be performed for the following reasons:
w If the results of a performance test are not within the specifications.
w After the replacement of a part or component that affects electrical performance.
The adjustment procedures require access to the interior of the instrument and therefore should only be performed by qualified service personnel. There are voltages at many points in the instrument which can, if contacted, cause personal injury. Be extremely careful. Adjustments should be performed only by trained service personnel.
Power is still applied to this instrument with the LINE switch in STANDBY. There is no OFF position on the LINE switch. Before removing or installing any assembly or printed circuit board, remove the power cord from the rear of both instruments and wait for the MAINS indicators (red LEDs) to go completely out.
Capacitors inside the instrument may still be charged even if the instrument has been disconnected from its source of power.
Use a
This manual contains procedures for testing and adjusting HP 8568B Spectrum Analyzers, including those with Option 001 (75 Ohm RF INPUT), Option 400 (400 Hz operation), Option 462, and Option 857 installed. The procedures in this manual can also be used to adjust HP 8568A Spectrum Analyzers that have been converted into HP 8568B Spectrum Analyzers through the installation of an HP 8568AB Retrofit Kit (formerly HP 8568A+ OlK Retrofit Kit).
G e n e r a l I n f o r m a t i o n
Operation Verification A high confidence level in the instrument???s operation can be achieved by running only the Operation Verification Program, since it tests most of the instrument???s specifications. It is recommended that the Operation Verification Program be used for incoming inspection and after repairs, since it requires much less time and test equipment.
A description of the program can be found in the Installation and Verification manual.
Option 462
Instruments
Option 857
Instruments
Option 462 instruments require that the performance tests and adjustment procedures listed below be performed instead of their standard versions included in chapters two and three. Information on Option 462 versions are located in Chapter 4, Option 462.
6 dB Bandwidths:
Test 4, 6 dB Resolution Bandwidth Accuracy Test
Test 5, 6 dB Resolution Selectivity Test
Adjustment 9, 6 dB Bandwidth Adjustments
Impulse Bandwidths:
Test 4, Impulse and Resolution Bandwidth Accuracy Test Test 5, Impulse and Resolution Selectivity Test
Test 6, Impulse and Resolution Bandwidth Switching Uncertainty Test
Adjustment 9, Impulse Bandwidth Adjustments
Option 857 instruments require that the performance test procedure listed below be performed instead of the standard version included in Chapter 2. Information on Option 857 is located in Chapter 5, Option
8 5 7 .
Test 12, Option 857 Amplitude Fidelity Test
G e n e r a l I n f o r m a t i o n 1 - 3
???Ihble
???Ihble
* Part of Service Accessories
G e n e r a l I n f o r m a t i o n
Figure
2
Performance Tksts
Introduction
The procedures in this section test the instrument???s electrical performance using the Specifications in the Installation and Verification Manual as the performance standards. None of the tests require access to the interior of the instrument. The manual
Performance Tests provided in this section should be performed only if
Verification of
Specifications
When a complete verification of specifications is required, proceed as follows:
1.Run the Operation Verification Program.
2.The Operation Verification Program verifies compliance with specifications of all tests it performs. The tests not performed by the Operation Verification Program must be done manually and are as follows:
nCenter Frequency Readout Accuracy
nSpurious Responses
H Fast Sweep Time Accuracy
n1st LO Output Amplitude Responses
nFrequency Reference Error
If the results of a performance test are marginally within specification, go to the Adjustments section of this manual and perform the related adjustment procedures. When an adjustment is directly related to a performance test, the adjustment procedure is referenced under RELATED ADJUSTMENT in the performance test.
Calibration Cycle This instrument requires periodic verification of performance. The instrument should have a complete verification of specifications at least every six months.
P e r f o r m a n c e T e s t s
Equipment Required Equipment required for the manual performance tests and adjustments is listed in ???Ihble
Test Record The Operation Verification Program provides a detailed test record when a printer is used with the controller. If manual performance tests are done, results of the performance tests may be tabulated in the HP 8568B Performance Test Record at the end of this section. The HP 8568B Performance Test Record lists all of the tested specifications and the acceptable ranges for the measurement values obtained during the tests.
Note Allow
able
2 - 2 P e r f o r m a n c e T e s t s
1. Center Frequency Readout Accuracy Test
1. Center
Frequency Readout
Accuracy ???I&t
Related Adjustments Frequency Control Adjustments
Time Base Adjustment
Step Gain and 18.4 MHz Local Oscillator Adjustments
50 MHz
Specification (uncorrected)
f2% of frequency span + frequency reference error x tune frequency +30% of resolution bandwidth setting + 10 Hz) in AUTO resolution bandwidth after adjusting FREQ ZERO at stabilized temperature.
Description A synthesized signal source that is
for several different frequency settings over the analyzer???s range. The signal source is
SYNTHESIZED
SHEEPER
SIONAL
INPUT 2
ADAPTER
CABLE ASSENBLY
Figure
P e r f o r m a n c e T e s t s
1. Center Frequency Readout Accuracy Test
Equipment Synthesized Sweeper . . . . . . . . . HP8340A Frequency Standard . 10 MHz standard, accy within + 1 part in lOlo,
e.g. HP 5061A
Adapter, Type N (m) to SMA (f) . . .
Procedure 1. Connect CAL OUTPUT to SIGNAL INPUT 2.
2 . Press [INSTR PRESET), @ on the analyzer.
3 . Adjust FREQ ZERO for a maximum amplitude trace.
4 . Press
5 . Set the synthesized sweeper for a 100.000 MHz signal at a level of approximately 0 dBm.
6 . Connect equipment as shown in Figure
7 . Set analyzer ~CENTER FREQUENCY) and [FREQUENCY SPAN) and synthesized sweeper frequency according to Table
8.Record the CENTER readout frequency in the table for each setting. The limits for this frequency are given in the table. See Figure
Figure
2 - 4 Performance Tests
Performance Tests 2 - 5
2. Frequency Span Accuracy lkst
Related Adjustments Frequency Control Adjustments
50 MHz
Specification
SpanUncertainty
>l MHz f(2% of the actual frequency separation between two points +0.5% of span setting)
51 MHz 445% of the actual frequency separation between two points +0.5% of span setting)
Description Frequency Span accuracy is determined by measuring a frequency at 5% of sweep and then at 95% of sweep. These frequencies correspond
to half a division from each edge of the CRT.
The spans chosen are based on the architecture of the HP 8568B RF hardware:
SYNTHESIZED WEEPER
Figure
Procedure 1. Connect equipment as shown in Figure
2.Press QNsTR PRESET] on analyzer.
3.Press [CENTER FREQUENCY] 100 MHz, [FREQUENCY SPAN_) 200 Hz.
4.Connect synthesized sweeper tot spectrum analyzer RF input 2.
5.On synthesized sweeper, select external REFERENCE and key in (jj??? 0 dBm.
6.Press Icw] and key in 99.999 910 MHz.
7.Press MARKER [PEAK SEARCH) on spectrum analyzer and record marker reading under FREQ C of Table
8.Set synthesized sweeper frequency to 100.000 090 MHz.
9.Press MARKER [PEAK SEARCH] and record marker reading under FREQ D of Table
10.Repeat the span measurement procedure of steps 6 through 9 for each frequency span listed in Table
11.Determine the frequency difference between the two measured points. Enter this value under the A DUT column in ???fable
12.The frequency span error is the difference between A DUT and A SYNTH. (See table
13.Compare the table
lhble
I
I
I
I
Performance Tests 2 - 7
2. Frequency Span Accuracy Test
For spans > 1 MHz, the spec is: >*[(.02)(A synth freq) + (.005)(span)]
For spans 2 1 MHz, the spec is: >&[(.05)(A synth freq) + (.005)(span)]
2 - 8 Performance Tests
3. Sweep Time Accuracy Test (220 ms)
3. Sweep Time Accuracy Test (220
ms)
Related Adjustment Frequency Control Adjustments
Specification *lo% for sweep times 5100 seconds &20% for sweep times >lOO seconds
Description Preferred Procedure
This test is for sweep times 220 ms. For faster sweep times, refer to
Fast Sweep Time Accuracy Test (Test 19).
A universal counter is connected to the PENLIFT RECORDER OUTPUT (on the rear panel) of the spectrum analyzer. The counter is used in time interval mode to determine the ???pen down??? (sweep time) interval of the PENLIFT RECORDER OUTPUT. The penlift output voltage level corresponds directly to the sweeping of the analyzer (pen down = OV) and
Alternate Procedure
Perform this procedure if the equipment for the preferred procedure is unavailable.
Sweep time accuracy for sweep times 220 ms can also be measured using the HP 8568B???s internal frequency counter for a time interval measurement.
SPECTRUM ANALYZER
UNIVERSAL
COUNTER
DIGITAL VOLTMETER
\
Figure
Performance Tests 2 - 9
3. Sweep Time Accuracy Test (220 ms)
Equipment Universal Counter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . HP 5316A Digital Voltmeter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . HP 3456A
Procedure
Sweep Times 220 ms 1. Connect equipment as shown in Figure
2.Press [INSTR PRESET) on the spectrum analyzer.
3.Key in the following settings:
4.Set up the universal counter as follows:
a. Set all front panel keys in ???out??? position. b. Set POWER switch to ON.
c.Set GATE TIME vernier control to 9 o???clock. d. Set SEPXOM A switch to COM A position.
e. Depress T.I. A + B switch (making sure the blue shift key is out).
f. Set Channel A trigger level to trigger on negative slope. g. Set Channel B trigger level to trigger on positive slope. h. Set both Channel A and Channel B ac/dc switches to de.
i.Connect the digital voltmeter to Channel A TRIGGER LEVEL OUT. (Be sure to ground the DVM properly.)
j. Adjust Channel A trigger level to set a DVM voltage reading of 0.3 v.
k. Repeat steps i and j for Channel B.
5.Set analyzer @WEEP TIME) to 20 ms. Allow the universal counter enough time to settle at this sweep time.
3. Sweep Time Accuracy Test (220 ms)
NOTE:PULSE WIDTH APPROXIMATE
Figure
6.Note the measured sweep time on the universal counter and record this value in Table
7.Repeat steps 5 and 6 for each sweep time setting in Table
???Ihble
Sweep Time Accuracy, Sweep Times 220 ms
8.Press MARKER (NORMAL].
9.Use @J to place the marker at the second vertical graticule. 10. Press @FiHIFT),[Xi!FF~.
11.Set analyzer [SWEEP TIME) to 20 s. Allow the universal counter enough time to settle at this sweep time.
12.Note the measured sweep time on the universal counter and
record this value in Table
13. Repeat steps 11 and 12 for 200 s sweep time.
P e r f o r m a n c e T e s t s
3. Sweep Time Accuracy Test (220 ms)
???Ihble
Sweep Time Accuracy, Sweep Times 220 s
Sweep Times 220 ms (Alternate Procedure)
Measurement
Sweep Time
Measurement
14.Sweep times 220 ms are tested without external test equipment by the following procedure.
15.Press ~NSTR PRESET).
16.Set @WEEP TIME] according to Table
17.Press [SHIFT) [REsy three times. The Active Function Block reads SWEEP GEN followed by a measured sweep time. This is the
18.Press MARKER (OFF).
19.Press @?i] (REST three times and note the SWEEP GEN reading. The limits for the SWEEP GEN reading are listed in Table
20.Repeat steps 16 to 19 for each sweep time shown in Table
lhble
Sweep Time Accuracy, Sweep Times 220 ms (Alternate Procedure)
2 - 1 2 P e r f o r m a n c e T e s t s
4. Resolution
Bandwidth
Accuracy Test
4. Resolution Bandwidth Accuracy Test
(For instruments with Option 462, refer to Chapter 4.)
Related Adjustment
Specification &20%, 3 MHz
+lO%, 3 kHz to 1 MHz
&20% 10 Hz to 1 kHz
30 kHz and 100 kHz bandwidth accuracy figures apply only with 190% Relative Humidity, < 40??C.
Description The 3 dB bandwidth for each resolution bandwidth setting is measured with the MARKER function to determine bandwidth
accuracy. The CAL OUTPUT is used for a stable signal source.
Equipment None Required
Procedure 1. Press [INSTR PRESET).
2.Connect CAL OUTPUT to SIGNAL INPUT 2.
3.Key in spectrum analyzer setting as follows:
(CENTER FREQUENCY) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20 MHZ
4.Press SCALE LIN pushbutton. Press csHIFT],[my (resolution bandwidth).
5.Adjust [REFERENCE LEVEL] to position peak of signal trace at reference level (top) graticule line. Press SWEEP [SINGLE).
6.Press MARKER [ml and place marker at peak of signal trace with DATA knob. Press MARKER In] and position movable marker 3 dB down from the stationary marker on the
7.Press MARKER (ZJ and position movable marker 3 dB down from the signal peak on the negative going edge of the trace (the MARKER A amplitude readout should be .OO dB f0.05 dB). The 3 dB bandwidth is given by the MARKER A frequency readout (see Figure
P e r f o r m a n c e T e s t s
4. Resolution Bandwidth Accuracy Test
Figure
8.Vary spectrum analyzer settings according to ???fable
???Ihble
2 - 1 4 P e r f o r m a n c e T e s t s
5.Resolution Bandwidth Selectivity Test
6.Read the 60 dB bandwidth for the 3 MHz resolution bandwidth setting from the MARKER A frequency readout (see Figure
7.Vary spectrum analyzer settings according to Table
procedure of steps 4 through 6 and record the value in ???fable
8.Record the 3 dB bandwidths from Table
9.Calculate the bandwidth selectivity for each setting by dividing the 60 dB bandwidth by the 3 dB bandwidth. The bandwidth ratios should be less than the maximum values shown in
???fable
10.The 60 dB bandwidth for the 10 Hz resolution bandwidth setting should be less than 100 Hz.
Figure
2 - 1 6 P e r f o r m a n c e T e s t s
5. Resolution Bandwidth Selectivity Test
???lhble
P e r f o r m a n c e T e s t s 2 - 1 7
6. Resolution
Bandwidth
Switching
Uncertainty Test
(For instruments with Option 462, refer to Chapter 4.)
Related Adjustments 3 MHz Bandwidth Filter Adjustments
21.4 MHz Bandwidth Filter Adjustments Down/Up Converter
Adjustments
Specification (uncorrected; referenced to 1 MHz bandwidth; 20 - 30??C after 1 hour
f0.8 dB, 30 Hz bandwidth *0.5 dB, 100 Hz to 1 MHz bandwidth
fl.O dB, 3 MHz bandwidth 30 kHz and 100 kHz bandwidth switching uncertainty figures only applicable 190% Relative Humidity
Description The CAL OUTPUT signal is applied to the input of the spectrum analyzer. The deviation in peak amplitude of the signal trace is then measured as each resolution bandwidth filter is switched in.
Equipment None Required
4 . Press LOG (ENTER dB/bIvj and key in 1 dB. Press MARKER
[PEAK SEARCH) a.
5 . Press m,@J
6 . Key in settings according to Table
[PEAK SEARCH] at each setting, then read the amplitude deviation from the MARKER A readout at the upper right of the display (see Figure
2 - 1 8 P e r f o r m a n c e T e s t s
6. Resolution Bandwidth Switching Uncertainty Test
Figure
???Ihble
Performance Tests
7.Input Attenuator Switching Uncertainty Test
6.Press MARKER CPEAK SEARCH),(KJ
7.Set [~~J???JREFERENCE LEVEL], and frequency synthesizer
amplitude according to Table
HKR A 0 Hz
Performance Tests 2 - 21
8. Frequency
Response Test
Related Adjustment Slope Compensation Adjustment
Specification SIGNAL INPUT 1
~tl.5 dB, 100 Hz to 1.5 GHz
*I dB, 100 Hz to 500 MHz
SIGNAL INPUT 2
fl dB, 100 kHz to 1.5 GHz
Description Frequency response at both analyzer inputs is tested by slowly sweeping a flat signal source over the frequency range and observing the
with a
SPECTRUM ANALYZER
SYNTHESIZED SWEEPER
POWER METER
R E C O R D E R
O U T P U T
I G N A L
N P U T
.\ I
ADA,PT,ERS
POWER SENSOR
O P T I O N 0 0 1 . A D D 5 0 OHMS/75 O H M P A D A N D A D A P T E R
gb12b
Figure
2 - 22 Performance Tests
8. Frequency Response Test
Note Equipment listed is for three test setups, Figure
Equipment Synthesized Sweeper . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . HP 8340A Power Meter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ..HP436A Power Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . HP 8482A Frequency Synthesizer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . HP 3335A Function Generator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . HP 3312A Power Splitter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . HP 11667A Adapter, Type N (m) to BNC (f) . . . . . . . . . . . . . . . . . . . . . . . . HP
Additional Equipment for Option 001:
5OfY700 Minimum Loss Pad . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . HP 11852A Adapter, Type N (f) to BNC (m) (7561) . . . . . . . . . . . . . . . . . . HP
Procedure
20 MHz to 1.5 GHz 1. Press ~NSTR PRESET) on spectrum analyzer and synthesized sweeper.
3.Connect equipment as shown in Figure
4.Depress RANGE HOLD button on power meter.
5.Select METER leveling on synthesized sweeper.
6.Key in the following spectrum analyzer settings:
[CENTER FREQUENCY) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .20 MHZ
P e r f o r m a n c e T e s t s 2.23
8. Frequency Response Test
7.Adjust POWER LEVEL on synthesized sweeper (using data knob) to place peak of 20 MHz signal near reference level (top) graticule line.
8.Press [ENTER dB/mv], 1 dB on spectrum analyzer. Adjust POWER LEVEL on synthesized sweeper to position peak of signal 2 divisions below the reference level line.
9 . Key in the following spectrum analyzer settings:
START FREQ) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20.MHz
10.Press TRACE A [MAX) on the analyzer.
11.Press SWEEP SINGLE on the synthesized sweeper.
DL
dBm
,I
Figure
12.Press DISPLAY LINE [ENTER) on the spectrum analyzer. Use the Display Line to measure the maximum and minimum points on the trace. Record measurements below.
SIGNAL INPUT 2
(20 MHz to 1.5 GHz)
MaximumdBm
MinimumdBm
13.To check SIGNAL INPUT 1, use the type N male to BNC male adapter to connect the power splitter directly to SIGNAL INPUT 1.
Option 001: Use HP 11852A Minimum Loss Pad and adapters between splitter and spectrum analyzer input.
14. Press [INSTR PRESET) on spectrum analyzer, then activate SIGNAL INPUT 1 with the pushbutton.
2 - 24 Performance Tests
8. Frequency Response Test
Option 001: Set [REFERENCE LEVEL] TO
15.Repeat steps 6 through 11. Press DISPLAY LINE [ENTER) on the spectrum analyzer. Use the Display Line to measure the maximum and minimum points on the trace. Record measurements below.
SIGNAL INPUT 1
(20 MHz to 1.5 GHz)
MaximumdBm
MinimumdBm
16.Press MARKER
SIGNAL INPUT 1
(20 MHz to 500 GHz)
MaximumdBm
MinimumdBm
100 kHz to 20 MHz 17. Set the frequency synthesizer controls as follows:
FREQUENCY . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20 MHz
SWEEP WIDTH . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19.9 MHz
AMPLITUDE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . - 2 dBm
(Option 001: + 4 dBm)
18.Connect equipment as shown in Figure
Option 001: Use HP 11852 Minimum Loss Pad and adapters.
SPECTRUM ANALYZER
FREQUENCY SYNTHESIZER
II
OPTIOW 001: MO 80 oHMa/ 78 ON88 PM AW AIDAPTUII
Figure
19. Press ~NSTR PRESET) on the spectrum analyzer. Activate SIGNAL INPUT 1 with the pushbutton.
20. Key in the following spectrum analyzer settings:
Performance Tests 2 - 25
8. Frequency Response Test
25.After completion of sweep, press DISPLAY LINE (j?KiK] on the spectrum analyzer. Use the Display Line to measure the maximum and minimum points on the trace. Record the measurements below.
SIGNAL INPUT 1 (100 kHz to 20 MHz)
MaximumdBm
MinimumdBm
26. Measure and record signal level at start of trace (100 kHz).
SIGNAL INPUT 1 (100 kHz)
dBm
27.Connect output of frequency synthesizer to SIGNAL INPUT 2. Activate this input with the pushbutton.
Option 001. Do not use HP 11852A Minimum Loss Pad. Set frequency synthesizer output amplitude to
28.Press TRACE A
29.Set frequency synthesizer FREQUENCY to 20 MHz. Set spectrum
analyzer (CENTER FREQUENCY_) to 20 MHz, and (FREQUENCY SPAN) to
1MHz.
30.Repeat steps 22 through 24.
31.After completion of sweep, press DISPLAY LINE [ENTER] on the spectrum analyzer. Use the Display Line to measure the maximum and minimum points on the trace. Record the measurements below.
SIGNAL INPUT 2
(100 kHz to 20 MHz)
MaximumdBm
MinimumdBm
Figure
Frequency Response Test Setup (100 Hz to 100 k=)
P e r f o r m a n c e T e s t s
8. Frequency Response Test
100Hzto 100 kHz 32. Press (INSTR PRESET) on the spectrum analyzer. Activate SIGNAL INPUT 1.
33. Key in the following spectrum analyzer settings:
34.Connect equipment as shown in Figure
35.Set the function generator controls as follows:
LINE . . . . . . . . . . . . . . . . . . .. . . . . . .. . . . . . . . . . . . . . . . . . . . ON
RANGE Hz . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 K
FUNCTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . -
OFFSET . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . CAL (button in)
AMPLITUDE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 V
AMPLITUDE VERNIER . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . midrange
SYM . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . CAL
TRIGGER PHASE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . FREE RUN
MODULATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . all out
MODULATION RANGE Hz . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . I
M O D U L A T I O N R A N G E H z V E R N I E R . . . . . . . f u l l y C C W
MODULATION SYM . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . CAL
Percent Modulation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .. ..fullyCW
36.Adjust function generator FREQUENCY to place signal between the last two graticule lines (right side) on the signal analyzer display.
37.Adjust AMPLITUDE VERNIER on the function generator until the peak of the signal is at the reference graticule line on the spectrum analyzer display.
38.Press LOG CENTER dB/DIv_) 1 dB on the spectrum analyzer. Press DISPLAY LINE [ENTER] and set the Display Line to the level recorded for 100 kHz in step 25.
39.Adjust function generator AMPLITUDE VERNIER to place peak of signal at the Display Line.
40.Adjust FREQUENCY on the function generator to position the signal trace at the right edge of the spectrum analyzer display (last graticule line).
41.Press MODULATION SWP on the function generator and allow the function generator to make at least two complete sweeps. Press
TRACE A [MAX HOLD). Allow the function generator to make one complete sweep. After completion of the sweep, press TRACE A
m).
2.28 P e r f o r m a n c e T e s t s
8. Frequency Response Test
42.Press DISPLAY LINE [ENTER] on the spectrum analyzer. Use the Display Line to measure the maximum and minimum points on the trace. (Disregard LO Feedthrough at 1 kHz.) Record the measurements below.
SIGNAL INPUT 1
(1 kHz to 100 kHz)
MaximumdBm
MinimumdBm
43.Set Display Line to peak of trace at 1 kHz.
44.Key in the following spectrum analyzer settings:
TRACE A
. . . . . . . . . . . . . . . . . . . . .l kHz 1 kHz iO0 Hz
45. Set function generator controls as follows:
RANGE Hz . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 100 (button) FREQUENCY . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 MODULATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ..all out
46.Adjust function generator FREQUENCY as necessary to place signal near center graticule line and adjust AMPLITUDE VERNIER to place peak of signal at Display Line.
47.Key in the following spectrum analyzer settings:
FREQUENCY SPAN) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 100 Hz
&TT!, . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30 Hz
48. Set (CF STEP SIZE] to 100 Hz. Step spectrum analyzer
( C E N T E R FFIEQUENCY] from 1 kHz to 100 Hz with a, while setting function generator FREQUENCY to match spectrum analyzer center frequency at each step. Record
P e r f o r m a n c e T e s t s 2 - 2 9
8. Frequency Response Test
49.For each input, subtract the lowest minimum level (greatest negative) from the highest maximum (least negative) measurement recorded in steps indicated. The result should not exceed 2 dB.
SIGNAL INPUT 1
100 Hz to 500 MHz (from steps 16, 25, 42, or 48)
SIGNAL INPUT 2
100 kHz to 1.5 GHz (from steps 12 or 31)
50.Subtract the lowest minimum level (greatest negative) from the highest maximum (least negative) measurement recorded in steps indicated. The result should not exceed 3 dB.
SIGNAL INPUT 1
100 Hz to 1.5 GHz (from steps 15, 16, 25, 42, or 48)
2.30 P e r f o r m a n c e T e s t s
9. RF Gain Uncertainty Test
h
Figure
2 - 3 2 P e r f o r m a n c e T e s t s
10. IF Gain Uncertainty Test
10. IF Gain
Uncertainty Test
Related Adjustments Step Gain and 18.4 MHz Local Oscillator Adjustments
21.4 MHz Bandwidth Filter Adjustments
Specification Assuming the internal calibration signal is used to calibrate the reference level at
Description The IF gain steps are tested over the entire range from 0 dBm to
Figure
Performance Tests 2 - 33
10. IF Gain Uncertainty Test
Equipment Frequency Synthesizer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . HP 3335A Adapter, Type N (m) to BNC (f) . . . . . . . . . . . . . . . . . . . . . . . . HP
Procedure 1. Press ( INSTR
2.Connect CAL OUTPUT to SIGNAL INPUT.
3.Press Cm] 8. Adjust AMPTD CAL for a MARKER amplitude of
4.Press (1~~733 PRESET).
10 dB Gain Steps 5. Set the frequency synthesizer for an output frequency of 20.0010 MHz and an output power level of
6.Connect the equipment as shown in Figure
7.Key in analyzer settings as follows:
CCENTER FREQUENCY) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20.001 MHZ
CFREQUENCY SPAN) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 kHz
8 Press MARKER (PEAK SEARCH).~~) or adjust (CENTER FREQUENCY) to center signal trace on display.
10.Press MARKER [PEAK SEARCH),
11.Press (SHIFT),(m]I to permit extended reference level settings.
12.Set the analyzer (REFERENCE LEVEL],(VIDEO], and frequency synthesizer amplitude according to Table
Note After measurement at the (REFERENCE LEVEL) =
2 - 34 Performance Tests
10. IF Gain Uncertainty Test
???Ihble
Figure
Performance Tests 2 - 35
10. IF Gain Uncertainty Test
2 dB Gain Steps 13. Press QNSTR pREsETj,(jRECALL) 7.
14.Set [REFERENCE LEVEL] to
15.Press MARKER (OFF). Set CVlDEo] to 100 Hz.
16.Set the frequency synthesizer for an output power level of
17.Press MARKER [PEAK SEARCH),@
18.Set the analyzer CREFERENCE LEVEL) and the frequency synthesizer amplitude according to ???Iable
???lhble
LREFERENCE LEVEL) Frequency Deviation
CdBm) Synthesizer (MARKER A
Amplitude ???Amplitude
YKRAE Hz
1 m/
Figure
2 - 36 Performance Tests
10. IF Gain Uncertainty Test
0.1dB Gain Steps 19. Set [REFERENCE LEVEL) to 0 dB.
20.Set the frequency synthesizer for an output power level of
21.Press MARKER [PEAK SEARCH),@.
22.Set the analyzer and the frequency synthesizer amplitude according to lkble
23.Find the largest positive deviation and the largest negative
deviation for reference level settings from 0 dBm to
AB
Reference Level Range: 0 to
24.Find the largest positive and negative deviations in ???lkble
Performance Tests 2 - 37
10. IF Gain Uncertainty Test
???Ihble
25.The sum of the positive deviations recorded in A, C, and D should not exceed 0.6 dB.
26.The sum of the negative deviations recorded in A, C, and D should not be less than
27.The sum of the positive deviations recorded in A, B, C, and D should not exceed 1.0 dB.
28.The sum of the negative deviations recorded in A, B, C, and D should not exceed - 1 .O dB.
2 . 3 8 P e r f o r m a n c e T e s t s
11. Log Scale Switching Uncertainty Test
b
2 dB/
Figure
able
Incremental
ho.1 dB/dB over 0 to 80 dB display Cumulative
3 MHz to 30 Hz Resolution Bandwidth
<kl.O dB max over 0 to 80 dB display (20 - 30??C). 5% 1.5 dB max over 0 to 90 dB display
Linear:
f3% of Reference Level for top
Description Amplitude fidelity in log and linear modes is tested by decreasing the signal level to the spectrum analyzer in 10 dB steps with a calibrated signal source and measuring the displayed amplitude change with the analyzer???s MARKER A function.
SIGNAL ANALYZER
SYNTHESIZERf LEVEL
GENERA OR
ADAPTER
?
Figure
P e r f o r m a n c e T e s t s
12. Amplitude Fidelity Test
Equipment Frequency Synthesizer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . HP 3335A Adapter, Type N (m) to BNC (f) . . . . . . . . . . . . . . . . . . . . . HP
Procedure Log Fidelity
1. Set the frequency synthesizer for an output frequency of 20.000 MHz and an output power level of + 10 dBm. Set the amplitude increment for 10 dB steps.
2.Connect equipment as shown in Figure
3.Press ~NSTR PRESET) on the analyzer. Key in analyzer settings as follows:
[ C E N T E R F R E Q U E N C Y ) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .20 MHZ FREQUENCY S P A N ] . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .50 kHz
CREFERENCE LEVEL) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . + 10 dBm
4.Press MARKER SPEAK SEARCHJ@~K~GTJ~MKR + REF LVL) to
center the signal on the display.
5.Key in the following analyzer settings:
FREQUENCY SPAN) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . O Hz
[&Xi, . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 Hz
6.Press MARKER A. Step the frequency synthesizer output amplitude from + 10 dBm to
recording it in column 2 of ???fable
7.Subtract the value in column 1 from the value in column 2 for each setting to find the fidelity error.
2 - 42 Performance Tests
12. Amplitude Fidelity Test
8.The fidelity error for amplitude steps from
9.The fidelity error at the
YKR A
REF 9.7 d&n ATTEN 20 dB
4J
10 d0/
Figure
Linear Fidelity 10. Key in analyzer settings as follows:
FREQUENCY SPAN) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .I MHz
[M, . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 MHz
11.Set the frequency synthesizer for an output power level of + 10 dBm.
12.Press SCALE LIN pushbutton. Press MARKER
CPEAK SEARCHJ[MKR) to center the signal on the display.
13.Set (FREQUENCY SPAN) to 0 Hz and [VlDEo] to 1 Hz. Press
14.Decrease frequency synthesizer output amplitude by 10 dB steps, noting the MARKER A amplitude and recording it in column 2 of
???Ikble
Performance Tests 2 - 43
12. Amplitude Fidelity Test
???Ihble
2 - 44 Performance Tests
13. Average Noise Level Test
13. Average Noise
Level Tkst
Specification
Option 001:
Description The average noise level is checked by observing the displayed noise level at several frequencies with no input signal applied.
Equipment 50 Ohm Termination . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .HP 11593A
Procedure 1. Press (~NsTR PRESET].
2.Connect CAL OUTPUT to SIGNAL INPUT 2.
3.Press m @. Adjust AMPTD CAL for a MARKER amplitude of
4.Press ~NSTR PRESET].
5.Disconnect CAL OUTPUT from analyzer. Terminate SIGNAL INPUT 2 with a 509 coaxial termination.
6.Key in spectrum analyzer settings as follows:
7.Press SWEEP (SINGLE] and wait for completion of the sweep.
8.Press DISPLAY LINE [ENTER). Using DATA knob, place display line at the apparent average amplitude of the noise trace (see Figure
Performance Tests 2 - 45
13. Average Noise Level Test
Figure
9.Read the average noise level from the DISPLAY LINE readout. The value should be
dBm
10.Change [CENTER FREQUENCY) to 1.001 MHz. Follow the procedure to steps 7 through 9 to determine the average noise level. The value should be
dBm
11.Change
dBm
2 - 46 Performance Tests
14. Residual Responses Test
14. Residual
Responses Test
Specification
Option 400:
Description The spectrum analyzer is checked for residual responses across its frequency range with no signal applied to the input and 0 dB input attenuation.
Equipment 50 Ohm Termination . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . HP 11593A
Procedure 1. Press (INSTR PRESET].
2.Connect CAL OUTPUT to SIGNAL INPUT 2.
3.Press IRECALL) @J Adjust AMPTD CAL for a MARKER amplitude of
4.Press QNSTR PRESET).
5.Disconnect CAL OUTPUT from analyzer. Terminate SIGNAL INPUT 2 with a 50 ohm coaxial termination.
6.Key in control settings as follows:
7.Press DISPLAY LINE CENTER_) and key in
8.Reduce [RESBW) or [VlDEOBW), if necessary, for a margin of at least 4 dB between the noise trace and the display line (refer to Figure
Note This test will require approximately 30 minutes to complete using the settings given in step 6. If the resolution bandwidth or video bandwidth are further reduced, a full band check of residual responses will take up to 15 hours to complete
Performance Tests 2 - 47
14. Residual Responses Test
Figure
9.Press SWEEP [SINGLE_) and wait for completion of sweep. Look for any residual responses at or above the display line. If a residual is suspected, press SWEEP CRINGLE) again and see if the response persists. A residual will persist on repeated sweeps, but a noise peak will not. Any residual responses must be
Option 400:
Any residual 500 Hz to 2.5 kHz must be
10.If a response appears marginal, do the following to determine whether or not it exceeds the specification.
a.Press ISAVE_) 0.
b.Press MARKER [NORMAL) and place the marker on the peak of the response in question.
c.Press MARKER Ijj), then activate SWEEP m.
d.Reduce [FREQUENCY SPAN] to 1 MHz or less. The amplitude of the response should be
e.Press (RECALL] (iJ to resume the search for residuals.
11.Step [CENTER FREQUENCY) to 1510 MHz with @) checking for residual responses at each step by the procedure of steps 9 and
10.There should be no residual responses at or above the display line below 1500 MHz.
Maximum Residual Response
dBm
MHz
2 - 46 Performance Tests
15. Spurious Responses Test
15. Spurious
Responses Test
Related Adjustment Second Converter Adjustments
Specification For total signal power of
Second Harmonic Distortion
For a signal
Third Order Intermodulation Distortion
For two signals each
Description Harmonic distortion (second and third) is tested using a signal source and a
Spurious responses due to image frequencies,
P e r f o r m a n c e T e s t s
15. Spurious Responses Test
SPECTRUM ANALYZER
SYNTHESIZED SMEEPEA
10 DB
ATTENUATOR
300 MHZ LPF
J
Figure
Note Equipment listed is for two test setups, Figure
Procedure Harmonic Distortion
1.Connect equipment as shown in Figure
2.On the spectrum analyzer, press [INSTR PRESET). Set the controls of the spectrum analyzer as follows:
3. On the synthesized sweeper, key in ~NSTR PRESET], Icw) 280 MHz,
( P O W E R L E V E L ) - 10 dBm.
4.On the spectrum analyzer, key in DISPLAY LINE CENTER]
5.On the synthesized sweeper, press ~POWER LEVEL] and use the ENTRY knob to adjust the amplitude of the displayed 280 MHz
15. Spurious Responses Test
signal for a marker indication of
6 On the spectrum analyzer, key in MARKER @,
CCENTER FREQUENCY) 560 MHz, MARKER CPEAK SEARCH) to position a second marker on the peak of the second harmonic distortion product of the 280 MHz input signal. The response should be below the display line (>70 dB below the input signal level).
7.On the synthesized sweeper, key in ~POWER LEVEL) ato decrease the amplitude of the 280 MHz signal by 10 dB.
8.On the spectrum analyzer, key in MARKER loFF),
CCENTER FREQUENCY) 280 MHz, [REFERENCE LEVEL)
9.On the synthesized sweeper, press [POWER LEVEL] and use the ENTRY knob to adjust the amplitude of the displayed 280 MHz signal for a marker indication of
10.On the spectrum analyzer, key in MARKER a],
[CENTER FREQUENCY) 840 MHz, MARKER [PEAK SEARCH) to position a second marker on the peak of the third harmonic distortion product of the 280 MHz input signal. The response should be below the display line (>75 dB below the input signal level).
Performance Tests 2 - 51
15. Spurious Responses Test
Intermodulation
Distortion
SPECTRUM ANALYZER
ATTENUATORATTENUATOR
DIRECTIONAL
50 MHz LON
PASS FILTER
Figure
11.Connect equipment as shown in Figure
12.Set the controls of the spectrum analyzer as follows:
CENTER FREQUENCY . . . . . . . . . . . . . . . . . . . . . . . . . . . 29.5 MHz
FREQUENCY SPAN . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 MHz
REFERENCE LEVEL . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
DISPLAY LINE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . OFF
13.On the synthesized sweeper, key in [cw 30 MHz, [POWER LEVEL],
14.On the frequency synthesizer, key in [FREQUENCY) 29 MHz,
(AMPLITUDE)
15. Set the controls of the spectrum analyzer as follows:
CENTER FREQUENCY . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29 MHz
FREQUENCY SPAN . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 500 Hz
16.On the spectrum analyzer, key in DISPLAY LINE (ENTER)
17.On the frequency synthesizer, adjust the signal amplitude for a marker indication of
18.On the spectrum analyzer, key in [CENTER FREQUENCY) 30 MHz, MARKER [PEAK SEARCH) to position a marker on the peak of the displayed 30 MHz signal.
19.On the synthesized sweeper, adjust the signal power level for a marker indication of
2 - 5 2 P e r f o r m a n c e T e s t s
15. Spurious Responses Test
Note If unable to locate intermodulation distortion products, temporarily increase output power level of frequency synthesizer and synthesized sweeper by + 10 dB. Return the output power level of both
signal sources to the previous settings before making distortion measurements.
20. On the spectrum analyzer, key in MARKER [n),
[CENTER FREQUENCY) 31 MHz, MARKER [PEAK SEARCH) to position
a marker at the peak of the 31 MHz
TO1 Distortion (1 MHz separation @ 30 MHz) dBm
21.On the spectrum analyzer, key in [CENTER FREQUENCY) 28 MHz, MARKER [PEAK SEARCH] to position a marker at the peak of the 28 MHz
TO1 Distortion (1 MHz separation @ 30 MHz) dBm
Q
2- 5 0
-6 0
-7 0
f2 - fl
S E C O N D
O R D E R
fl fz
S E C O N D
H A R M O N I C S
F R O M S I G N A L
GENERAToRS\
211 2f2
-
L
S E C O N D
T H I R DO R D E R
O R D E R
Figure
22.On the frequency synthesizer, key in [FREQUENCY) 29.99 MHz.
23.On the spectrum analyzer, key in MARKER IOFF),
CCENTER FREQUENCY) 29.99 MHz, DISPLAY LINE [ENTER]
MARKER (PEAK SEARCH).
Performance Tests 2 - 53
15. Spurious Responses Test
24.On the frequency synthesizer, readjust the signal amplitude as necessary to position the peak of the displayed 29.99 MHz signal at the top CRT graticule line.
25.On the spectrum analyzer, key in MARKER @,
(CENTER FREQUENCY) 30.01 MHz, MARKER (PEAK SEARCH) to position a second marker at the peak of the 30.01 MHz
TO1 Distortion (10 kHz separation @ 30 MHz) dBm
26.On the spectrum analyzer, key in [CENTER FREQUENCY) 29.98 MHz, MARKER [PEAK SEARCH) to position a second marker at the peak of the 29.98 MHz
TO1 Distortion (10 kHz separation @ 30 MHz) dBm
27.On the synthesized sweeper, press (POWER LEVEL] and decrease the amplitude of the 30 MHz signal by 13.0 dB from the current setting.
28.On the frequency synthesizer, key in [FREQUENCY] 29 MHz, (AMPLITUDE] and then decrease the amplitude of the 29 MHz signal by 13.0 dB from the current setting.
29.Set the controls of the spectrum analyzer as follows:
CENTER FREQUENCY . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29 MHz
FREQUENCY SPAN . . . . . . . . . . . . . . . . . . . . . . . . . ,500 Hz
REFERENCE LEVEL . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
MARKER . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . OFF
30.On the spectrum analyzer, key in DISPLAY LINE [ml
31.On the frequency synthesizer, adjust the signal amplitude for a marker indication of
32.On the spectrum analyzer, key in CCENTER FREQUENCY) 30 MHz, MARKER [PEAK SEARCH) to position a marker on the peak of the displayed 30 MHz signal.
33.On the synthesized sweeper, adjust the signal power level for a marker indication of
34.On the spectrum analyzer, key in MARKER a,
[CENTER FREQUENCY) 1 MHZ, MARKER [PEAK SEARCH] to position
a second marker at the peak of the 1 MHz
SO1 Distortion (1 MHz separation @ 30 MHz) dBm
2 - 5 4 P e r f o r m a n c e T e s t s
15. Spurious Responses Test
35.On the spectrum analyzer, key in [CENTER FREQUENCY] 59 MHz, MARKER [PEAK SEARCH) to position a second marker at the peak of the 59 MHz
SO1 Distortion (1 MHz separation @ 30 MHz) dBm
Performance Tests 2 - 55
16. Residual FM
Test
Specification <3 Hz
Description The spectrum analyzer CAL OUTPUT is used to supply a stable 20 MHz signal to the analyzer. The analyzer is tuned in zero span to a point on the 30 Hz bandwidth response for which the slope of the response is known from direct measurement. The residual FM is then slope detected over a 10 second interval, yielding a trace whose
Equipment None Required
Procedure 1. Press (~N~TR PRESET).
2.Connect CAL OUTPUT to SIGNAL INPUT 2.
3.Press (RECALL) 8 and adjust AMPTD CAL for a MARKER amplitude of
4.Press CRECALL) 9 and adjust FREQ ZERO for a maximum amplitude trace.
5.Set PREFERENCE LEVEL) to
6.Set ~FREQUENCY SPAN) to 100 Hz. Press SWEEP CSINGLE) and wait for completion of the sweep.
7.Press MARKER Cm), and place marker 1 division above the center graticule line on the
2 - 5 6 P e r f o r m a n c e T e s t s
16. Residual FM Test
hr RF
Figure
8.Compute the detection slope of the 30 Hz filter between the markers by dividing the MARKER A amplitude by the MARKER A frequency:
filter slope = MARKER A amplitude/MARKER Afrequency = dB/Hz
9.Press SWEEP [CONT),(mj IOFF).
10.Change FREQUENCY SPAN] to 0 Hz. Readjust FREQ ZERO, if necessary, to position the trace at the center graticule line. The amplitude variations of the trace (see Figure
P e r f o r m a n c e T e s t s
16. Residual FM Test
Figure
11.Press SWEEP CRINGLE) and wait for completion of the sweep.
12.Press MARKER [PEAK SEARCH_). Press DISPLAY LINE (m??? and position the display line at the lowest point on the trace.
Figure
2 - 5 8 P e r f o r m a n c e T e s t s
16. Residual FM Test
13.Press MARKER Ia] and position movable marker at the lowest point on the trace (see Figure
14.Divide the
= residual FM
The residual FM should be less than 3 Hz.
15.Press ~NSTR PRESET].
16.Press Cm] 9 and adjust FREQ ZERO for a maximum amplitude trace.
Performance Tests 2 - 59
17.
Sidebands Tests
Specification
Description
Equipment
Procedure
95 dB below the peak of a CW signal. Option $00: >75 dB below the peak of a CW signal.
The spectrally pure calibrator signal of the spectrum analyzer is applied to the analyzer input and the line related sidebands near the signal are measured.
None required
1.Press ~NSTR PRESET) on the analyzer. Connect CAL OUTPUT to
SIGNAL INPUT 2.
2.Press
3.Press (~NSTR PRESET).
4.Key in the following analyzer settings:
5. Wait for completion of sweep, then press MARKER
[ P E A K SEARCH),(jj).
6.Press (SHIFT)
7.Press MARKER (PEAK SEARCH],(~) and position movable marker at the peak of each line related sideband (120 Hz, 180 Hz, and 240 Hz for 60 Hz line frequency; 100 Hz, 150 Hz, and 200 Hz for 50 Hz line frequency, etc.). The MARKER A amplitude for each sideband should be
17.
Figure
O p t i o n 4 0 0 1. Press ~NSTR PRESET]. Connect CAL OUTPUT to SIGNAL INPUT 2.
2.Press (ml 8 and adjust AMPTD CAL for a MARKER amplitude of
3.Press ONSTR PRESET).
4.Key in the following analyzer settings:
6.Press (SHIFT) [VlDEo] G, SWEEP
7.Press MARKER SPEAK SEARCH],@ and position movable marker at the peak of each line related sideband (400 Hz, 800 Hz, and 1200 Hz). The MARKER A amplitude for each sideband should be
Performance Tests 2 - 61
18. Calibrator
Amplitude
Accuracy Test
Related Adjustment 20 MHz Reference Adjustments
Specification
Description The output level of the calibrator signal is measured with a power meter.
SPECTRUW ANILYZER
Figure
Equipment Power Meter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . HP 436A Power Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . HP 8482A
Adapter, Type N (f) to BNC (m) . . . . . . . . . . . . . . . . . . . . . . . . HP
Procedure 1. Connect equipment as shown in Figure
2.Measure output level of the CAL OUTPUT signal. The value should be
dBm
2 - 62 Performance Tests
19. Fast Sweep Time Accuracy Test (~20 ms)
19. Fast Sweep
Time Accuracy
Test (430 ms)
Related Adjustment None
Specification &lo% for sweep times 5100 seconds
Description The triangular wave output of a function generator is used to modulate a 500 MHz signal which is applied to the spectrum analyzer SIGNAL INPUT. The signal is demodulated in the zero span mode to display the triangular waveform. Sweep time accuracy for sweep times ~20 ms is tested by checking the spacing of the signal peaks on the displayed waveform.
SPECTRUM ANALYZER
SYNTHESIZED SWEEPER
Procedure 1. Connect equipment as shown in Figure
2.Press QNSTR PRESET) on spectrum analyzer.
3.Key in analyzer settings as follows:
4.Set synthesized sweeper for an output frequency of 500 MHz and an output power level of
Performance Tests 2 - 63
19.Fast Sweep Time Accuracy Test (~20 ms)
5.Press MARKER (PEAK SEARCHJ@~EFZF),~.
6.Set [FREQUENCY SPAN) to 0 Hz,
7.Set synthesized sweeper for an
8.Set function generator controls as follows:
9.Key in (SWEEP TIME] 5 ms and set function generator for a counter reading of 2.00 f0.02 kHz.
10.Adjust spectrum analyzer TRIGGER LEVEL to place a peak of the triangular waveform on the first graticule from the left edge of the CRT display as a reference. (Adjust function generator amplitude, if necessary, to provide a signal large enough to produce a stable display). The fifth peak from the reference should be within ~l~0.5 division of the sixth graticule from the left edge of the display (see Figure
11.Using sweep times and function generator frequencies in
Table
Figure
2 - 64 Performance Tests
19. Fast Sweep Time Accuracy Test (~20 ms)
Performance Tests 2 - 65
20. 1st LO Output Amplitude Test
Specification >+4 dBm from 2.0 GHz to 3.7 GHz
Description The power level at the 1ST LO OUTPUT connected is measured as the first L.O. is swept over its 2.0 GHz to 3.1 GHz range.
SPECTRUM ANALYZER
POWER METER
Figure
Equipment Power Meter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . HP 436A Power Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . HP 8482A
Procedure 1. Press (INSTR PRESET].
2.Set [SWEEP TIME) to 100 seconds.
3.Calibrate power meter and sensor. Connect equipment as shown in Figure
4.Observe the meter indication as the analyzer makes a complete sweep. The indication should be > + 4 dBm across the full sweep range.
dBm
5. Replace 50 ohm terminator on 1ST LO OUTPUT.
2 - 6 6 P e r f o r m a n c e T e s t s
21. Frequency Reference Error Test
21. Frequency
Reference Error
Test
Related Adjustment Time Base Adjustment
Specification Aging Rate
<l x 10eg/day and ~2.5 x 10m7 year; attained after 30 days warmup from cold start at 25??C.
Temperature Stability
<7 x
Description The frequency of the spectrum analyzer time base oscillator is measured directly using a frequency counter locked to a frequency reference which has an aging rate less than
Note This test requires that the spectrum analyzer be turned on (not in STANDBY) for a period of 30 days to ensure that the frequency reference attains its aging rate. However, after aging rate is attained, the frequency reference typically attains aging rate again in 72 hours of operation after being off for a period not exceeding 24 hours.
Care must be taken not to disturb the spectrum analyzer during the
is returned to its original position. Frequency changes due to mechanical shock will usually appear as a fixed frequency error.
The frequency reference is also sensitive to temperature changes; for this reason the ambient temperature near the instrument at the first measurement time and the ambient temperature at the second measurement time should not differ by more than 1??C.
Placing the spectrum analyzer in STANDBY mode turns the instrument off while continuing to provide power for the frequency reference oven, helping to minimize warmup time. However, the frequency reference must be on to attain its aging rate.
Performance Tests 2 - 67
???lhble
Performance Tests
Tkst 1. Center
Frequency Readout
Accuracy Test
Step 8. Center Frequency Readout Error Test Record
Test 2. Frequency Span Accuracy Test
???I&t 2. Frequency
Span Accuracy Test
Steps 7, 9, and 11. Wide Span Error
Step 12. Span Error
Note The specification in Table
For spans > 1 MHz, the spec is: >&[(.02)(A synth freq) + (.005)(span)]
For spans 5 1 MHz, the spec is: >&[(.05)(A synth freq) + (.005)(span)]
Performance Tests
Test 3. Sweep
Time Accuracy
Step 6. Sweep Time Accuracy, Sweep Times 220 ms
Step 12. Sweep Time Accuracy, Sweep Times 220 s
21
Step 19. Sweep Time Accuracy, Sweep Times 220 ms (Alternate Procedure)
???I&t 4. Resolution
Bandwidth
Accuracy
Test 4. Resolution Bandwidth Accuracy
Performance Tests
Test 5. Resolution
Bandwidth
Selectivity
Steps 7, 8 and 9. Resolution Bandwidth Selectivity
Test 6. Resolution Bandwidth Switching Uncertainty Test
lkst 6. Resolution Bandwidth Switching Uncertainty Test
Step 6. Bandwidth Switching Uncertainty
Performance Tests
Test 7. Input
Attenuator
Switching
Uncertainty Test
Test 8. Frequency Response Test
Performance Tests
Test 9. RF Gain
Uncertainty Test
Step 6. 2nd LO Shift
Min Measured Max
Test 10. IF Gain Uncertainty Test
Test 10. IF Gain
Uncertainty Test
Step 12. Step IF Gain Uncertainty, 10 dB Steps
Step 18. IF Gain Uncertainty, 2 dB Steps
Performance Tests
Test 10. IF Gain Uncertainty Test
Step 22. IF Gain Uncertainty, 0.1 dB Steps
Step 24. Recorded deviations from Steps 18 and 22.
26.Sum of negative deviations of A, C, & D
28.Sum of negative deviations of A, B, C, & D
Performance Tests
YLkst 11. Log Scale
Switching
Uncertainty Tkst
Step 6. Log Scale Switching Uncertainty
SCALE MKR Amplitude
(dB/DIV) PW
1
2
5
10
Deviation Allowable
(W Deviation
(W
xto.5
hO.5
f0.5
Test 12. Amplitude Fidelity Test
Test 12. Amplitude
Fidelity ???I&t
Step 6. Log Amplitude Fidelity
Step 14. Linear Amplitude Fidelity
Performance Tests
Test 13. Average
Noise Level Test
Test 14. Residual Responses Test
Test 14. Residual
Responses Test
Step 11. Maximum Residual Response
Performance Tests
Tkst 15. Spurious
Responses Tkst
Test 16. Residual FM Test
Test 16. Residual
FM Test
Step 14. Residual FM
IMinIMeasuredIrax]
Performance Tests
Tkst 17.
Sidebands Test
Test 18. Calibrator Amplitude Accuracy Test
Test 18. Calibrator
Amplitude
Accuracy ???I&t
Step 2. CAL OUTPUT Amplitude
Performance Tests 2.89
Test 19. Fast
Sweep Time Accuracy Test (~20
ms)
Step 11. F&t Sweep Time Accuracy (~20 ms)
[SWEEP TIME) Function Generator Frequency
(kW
2.90 Performance Tests
Test 20. 1st LO Output Amplitude Test
Test 20. 1st LO Output Amplitude Test
Step 4. 1st LO Output Level
Performance Tests
Test 21. Frequency
Reference Error
Test
3
Adjustments
Introduction
Warning
The procedures in this section are for the adjustment of the instrument???s electrical performance characteristics.
The procedures require access to the interior of the instrument and therefore should only be performed by qualified service personnel. Refer to Safety Considerations in this introduction.
1. Low Voltage Power Supply
The adjustment procedures should not be performed as routine maintenance, but only when Performance Tests cannot meet specifications. Before attempting any adjustment, allow the instrument to warm up for one hour. ???lhble
Adjustments
Safety
Considerations
Warning
Equipment
Required
Adjustment Tools
Although this instrument has been designed in accordance with international safety standards, this manual contains information, cautions, and warnings which must be followed to ensure safe operations and to retain the instrument in safe condition. Service and adjustments should be performed only by qualified service personnel.
Adjustments in this section are performed with power supplied to the instrument while protective covers are removed. There are voltages at many points in the instrument which can,
if contacted, cause personal injury. Be extremely careful. Adjustment should be performed only by trained service personnel.
Power is still applied to this instrument with the LINE switch in STANDBY. There is no OFF position on the LINE switch. Before removing or installing any assembly or printed circuit board, remove the power cord from the rear of both instruments and wait for the MAINS indicators (red LEDs) to go completely out.
Capacitors inside the instrument may still be charged even if the instrument has been disconnected from its source of power.
Use a
The equipment required for the adjustment procedures is listed in ???lkble
listed in the table. The test setup used for an adjustment procedure is referenced in each procedure.
For adjustments requiring a
any adjustment control in the analyzer. This is especially critical when tuning variable
???able
Components
Adjustments
Related
Adjustments
Location of lkst Points and Adjustments
Any adjustments which interact with, or are related to, other adjustments are indicated in the adjustments procedures. It is important that adjustments so noted are performed in the order indicated to ensure that the instrument meets specifications.
Illustrations showing the locations of assemblies containing adjustments, and the location of those adjustments within the assemblies, are contained within the adjustment procedures to which they apply. Major assembly and component location illustrations are located at the rear of this manual.
For Serial Prefix 3001A and below, see back of table for exceptions to AlA through AlA6.
Adjustments
???able
Adjustments
Adjustments
???lhble
Adjustments
???Ihble
???Ihble
Function of Component
Sets intensity level. Sets intensity level. Sets intensity level. Sets intensity level.
Compensates for feedthrough of INTG signal to Ul.
Compensates for feedthrough of INTG signal to Ull.
Compensates for DAC ladder resistance. Compensates for DAC ladder resistance.
Sets adjustment range of A4AlR36 FS Compensates for ON resistance of A4AlQ6
Sets adjustment range of LG20.
Adjusts log fidelity.
Log fidelity.
Adjusts overall linear gain.
Sets adjustment range of ATTEN.
Temperature compensation
Temperature compensation
Temperature compensation
Temperature compensation
Temperature compensation
Temperature compensation
Adjusts bandpass filter shape in wide bandwidths (> 100 kHz).
Sets adjustment range of CTR. Sets adjustment range of CTR. Log fidelity
Log fidelity Log fidelity Log fidelity Log fidelity Log fidelity
Sets adjustment range of LGlO. Sets adjustment range of AMPTD.
Adjustments
l???hble
Adjustments
Adjustments
???able
???able
Adjustments
???lkble
rapaors
Value (pF) 1BP Fart Number CD Value (pF) ZP Fart Number s-
-
Adjustments
???Ihble
Standard Value Replacement 0.125 Resistors (continued)
Type:
Range: 10 to 464K Ohms
Wattage: 0.125 at 125??C
To1 e???an fl.O%
Value (0) BP Part Number co Value (!I) ElP Fart Number
-
Table
Resistors
Type:
Range: 10 to 1.47M Ohms
Wattage: 0.5 at 125???C
To1???ant 31.0%
EIP Fart Number 5 Value (fl) BP Part Number 56 Value (0)-
-
Adjustments
Ihble
Standard Value Replacement 0.5 Resistors
(continued)
Type:
Range: 10 to 1.47M Ohms
Wattage: 0.5 at 125??C
To1 e*an fl.O%
Value (0) EIP Fart Number CD Value (a) EIP Fart Number
-
1.
Power Supply
Adjustments
Reference
1.
AlA f15 V Regulator
AlA + 120 V, +5.2 V Regulator (Serial Number Prefix 3004A and above)
AlA + 100 V, +5.2 V Regulator (Serial Number Prefix 3001A and below)
RF Section:
A24 Voltage Regulator
Description The + 15 V supply is adjusted for the
SPECTRUM ANALYZER
DIGITAL VOLTMTER
ADAPTER
(+)
Equipment
Procedure
Note
Figure
Digital Voltmeter (DVM) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . HP 3456A
1.Position the instrument on its right side with the
2.Set the LINE switch to ON and press (Ip). Mains indicator AlABDSl (red LED) in the
Use Figure
3. Verify that the + 15 V indicator AlAGDSl (yellow LED) is lit.
Adjustments
1.
4.Connect the DVM to AlA6TP3 on the
indication should be + 15.000 fO.O1O V dc. If the voltage is out of tolerance, adjust AlA6R9 + 15 V ADJ for the specified voltage.
, AlA71P2
AlA6DS2
AlA6TP4 5
Figure
Below)
A l A7DS2
AlA7TP3
A lA6DS2
AiA6TP4
t
Figure
Above)
5.Verify that the
6.Connect the DVM to AlA6TP4. DVM indication should be
Note
Note
1.
7.Verify that the + 120 V indicator AlA7DS2 (yellow LED) is lit.
On
8.Connect the DVM to AlA7TP3. DVM indication should be + 120.0 ~k3.0 V dc. The + 120 V supply is referenced to the + 15 V supply; therefore, if the + 120 V supply is out of tolerance, a circuit malfunction is indicated.
On
9.Verify that the +5.2 V indicator AlA7DSl (yellow LED) is lit.
10.Connect the DVM to AlA7TP2. DVM indication should be +5.200 ho.050 V dc. The +5.2 V supply is referenced to the + 15 V supply; therefore, if the +5.2 V supply is out of tolerance, a circuit malfunction is indicated.
RF Section 11. The +2OV indicator A24DS2 (yellow LED) should be lit. See Figure
A24
V O L T A G EA24R60
R E G U L A T O R +2OV A D J
A24
Figure
12.Connect the DVM to A24TP3 with the ground lead to A24TPl. Adjust A24R60 +2OV ADJ for a DVM indication of +20.000 &O.OlO V dc.
13.The + 15V indicator A24DS4 (yellow LED) should be lit.
14.Connect the DVM to A24TP2. The DVM indication should be
+15.000 ho.050 V dc. The + 15V supply is referenced to the
+2OV supply, therefore, if the + 15V supply is out of tolerance, a circuit malfunction is indicated.
15.The +5V indicator A24DS5 (yellow LED) should be lit.
16.Connect the DVM to A24TP5. The DVM indication should be +5.230 f0.050 V dc. The +5V supply is referenced to the +2OV
Adjustments
1.
supply, therefore, if the +5V supply is out of tolerance, a circuit malfunction is indicated.
17.The
18.Connect the DVM to A24TP7. The DVM indication should be
19.The
20.Connect the DVM to A24TP4. The DVM indication should be
2.
Adjustment (SN
3001A and Below)
Note
2.
This procedure is for
Note This procedure should be performed whenever the AlAll High Voltage Multiplier, AlVl CRT, or AlA High Voltage Regulator Assembly is repaired or replaced.
Reference
AlA
AlA
AlA f15 V Regulator
AlA + 100 V, +5.2 V Regulator
Description
Warning This procedure is intended for adjustment purposes only. Voltages are present which, if contacted, could cause serious personal injury. Approximately
If for any reason the AlA High Voltage Assembly or the post accelerator cable must be removed, refer to ???Discharge Procedure for High Voltage and CRT??? at the end of this adjustment procedure.
A 1OOO:l divider probe is used to measure the CRT cathode voltage. First, the
Adjustments
2.
DIGITIZING
OSCILLOSCOPE
SIGNAL ANALYZER
Figure
Equipment Digital Voltmeter (DVM) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .HP 3456A DC
Adjustment Procedure
Warning
In the following procedure, it is necessary to probe voltages which, if contacted, could cause serious personal injury. Use a nonmetallic alignment tool when making adjustments. Be extremely careful.
Warning Do not attempt to measure the CRT filament voltage directly. The filament voltage is referenced to the
1.Set the spectrum analyzer???s LINE switch to STANDBY.
2.Remove the top cover from the
3.Set the DVM to the 100 V range, and connect the DVM to AlA7TP3 (+ 100 V). Do not use the
Figure
3.30 A d j u s t m e n t s
Note
Warning
2.
The accuracy of the
I
. AlA6R32
Figure
4.Set the LINE switch to ON. Set the
5.Note the DVM indication at AlA7TP3.
DVM Indication:
6.Connect the
7.Note the DVM indication.
DVM Indication:
8.Divide the DVM indication in step 7 by the DVM indication in
step 5. This gives the calibration factor needed to compensate for
Calibration Factor:
9.Disconnect the
The MAINS
Figure
Adjustments
2.
12.Connect the
AlA
H,gh V o l t a g e
R e g u lotor
Figure
13.Reconnect ac line cords to both instrument sections. Set the LINE switch to ON.
14.Wait approximately 30 seconds for the dc regulator circuits to stabilize.
15.Adjust AlA6R32 HV ADJ for a DVM indication equal to the calibration factor (calculated in step 8) times the voltage labeled on the top of AlA
V dc
EXAMPLE:
Focus and Intensity
Adjustments
2.
If the calibration factor calculated in step 8 is 0.00099, and AlA3Tl is labeled for
0.00099 x
16.With the
17.Readjust AlA6R32 HV ADJ for a DVM indication equal to the voltage determined in step 15.
18.If a new CRT has just been installed do the following:
a.Set the
b.Wait an additional 30 minutes for the CRT to normalize.
c.Readjust AlA6R32 HV ADJ for a DVM indication equal to the voltage determined in step 15.
19.Set the LINE switch to STANDBY. Remove the ac line cord from each instrument section.
20.Wait at least one minute for the MAINS
21.Disconnect the
22.Remove the A3A2 Intensity Control Assembly from the
HP part number
23.Connect a calibrated 1O:l divider probe to the oscilloscope Channel 1 input.
24.On the oscilloscope, press (RECALL) (CLEAR) to perform a soft reset.
25.On the oscilloscope, press CCHAN], more preset probe , select channel 1, and use the
26.Set the oscilloscope controls as follows:
Press [CHAN):
Channel 1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . on amplitude scale . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .lO.OV/div offset . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ..6O.OOOOV coupling . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ..dc
Press @%KiZi@:
time scale . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ..50~s/div Press ITRIG):
EDGE TRIGGER . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . auto, edge source . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ..l level . . . . . . . . . . . . . . . . . . . . . . . . . .75.0000 V, rising edge
Press [DISPLAY):
connect dots . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ..on
Adjustments
2.
27.On the oscilloscope press [SHOW].
28.Connect the oscilloscope channel 1 probe to AlA3TP5 using a long probe extension. See Figure
29.Reconnect the ac line cords to each instrument section. Adjust the
30.Wait approximately 30 seconds for the dc regulator circuits to stabilize again.
31.With the
Note The AlA2R35 INT LIMIT adjustment compensates for the variation in beam
Figure
32.Using a
33.Adjust AlA2R35 INT LIMIT until the dot just disappears.
2.
34.On the oscilloscope, adjust the channel 1 offset voltage as
necessary to measure the
1 1 0 . 0 V/div offset: 6 0 . 0 0 v
10.00 : 1 dc
1 f 7 5 . 0 0 v
Figure
35.Connect a separate function generator to each of the X and Y inputs of the Display Adjustment Board, as shown in Figure
Set the function generators as follows:
36.Adjust AlA2R35 INT LIMIT clockwise until the display is just visible.
37.Adjust AlA4R7 POS, AlA5R7 POS, and if necessary the function generator de offsets for a
38.Set the
Adjustments
2.
39.Slowly adjust the
If the
a.Set the INTENSITY control fully counter clockwise.
b.Set the LINE switch to STANDBY.
c.Increase the value of AlA2R9.
d.Return to step 34.
Note Maximum CRT life expectancy is obtained when the
5 0 . 0 uS/dlv
1 f 85.00 v
Figure
40.Replace the cover on the AlA
41.The
Discharge Procedure for High Voltage and
CRT
Warning
2.
AlA4, and AlA assemblies function properly and do not require compensation, proceed directly to adjustment procedure 4, ???Final Display Adjustments (SN 3001A and Below)???.
The adjustment procedures in this manual do not require the removal or discharge of the AlA
This procedure should be performed by qualified personnel only. Voltages are present which, if contacted, could cause serious personal injury. Approximately
for several days if the
Warning Do not handle the AlA
1.Set the spectrum analyzer???s LINE switch to STANDBY, remove the ac line cords, and remove the AlA High Voltage Regulator safety cover.
Warning With the ac power cord disconnected, voltages are still present which, if contacted, could cause serious personal injury.
Adjustments
2.
5.While holding the insulated handle of the screwdriver, touch the
grounded blade to the following connections:
a.Both brown wires going to the rear of the CRT from AlA via cable harness W21.
b.The yellow, blue, and orange wires in the same cable as ???a. ??? above.
c.The top lead of each of the 11 large vertical capacitors on the AlA
6.Connect the jumper wire from chassis ground to the black wire coming from the AlAll
AlA
H i g h V o l t a g e
,T 1
AlA3TlAIA3
Figure
7.Remove all jumper wires. The AlA
8.A small bracket and screw secure the AlA
2.
Adjustment (SN
3004A and Above)
Note
2.
This procedure is for
Note
Reference
This procedure should be performed whenever the AlVl CRT or AlA High Voltage Regulator Assembly is repaired or replaced.
A lA2
AlA
AlA &15 V Regulator
AlA + 120 V, +5.2 V Regulator
Description
Warning This procedure is intended for adjustment purposes only. Voltages are present which, if contacted, could cause serious personal injury. Approximately
If for any reason the AlA High Voltage Assembly or the post accelerator cable must be removed, refer to ???Discharge Procedure for High Voltage and CRT??? at the end of this adjustment procedure.
A 1OOO:l divider probe is used to measure the CRT cathode voltage. First, the
Adjustments
2.
Equipment Digital Voltmeter (DVM) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .HP 3456A DC
Adjustment Procedure
Warning In the following procedure, it is necessary to probe voltages which, if contacted, could cause serious personal injury. Use a nonmetallic alignment tool when making adjustments. Be extremely careful.
Warning Do not attempt to measure the CRT filament voltage directly. The filament voltage is referenced to the
1.Set the spectrum analyzer???s LINE switch to STANDBY.
2.Remove the top cover from the
SPECTRUM ANALYZER
Figure
3.Set the DVM to the 1OOV range, and connect the DVM to AlA7TP3 (+ 120V) without the
Note The accuracy of the
2.
/ AlABOSl
- AlA7TP3
AlA H I G H
V O L T A G E
/ R E G U L A T O R
- AlA6R103
Figure
4.Set the LINE switch to ON. Set the
5.Note the DVM indication at AlA7TP3.
DVM Indication:
6.Connect the
7.Note the DVM indication.
DVM Indication:
8.Divide the DVM indication in step 7 by the DVM indication in
step 5. This gives the calibration factor needed to compensate for
Calibration Factor:
9.Disconnect the
Warning The MAINS
Figure
Adjustments
2.
11.Remove the protective cover from the AlA
use in step 15. In cases where more than one voltage is listed on this label, record the value which is closest to
V dc
Warning With power supplied to the instrument, AlA3TP2A is at a voltage level of approximately
12.Connect the
AlA3TP2A
Figure
13.Reconnect ac line cords to both instrument sections. Set the LINE switch to ON.
14.Wait approximately 30 seconds for the de regulator circuits to stabilize.
15.Adjust AlA6R103 HV ADJ for a DVM indication equal to the calibration factor (calculated in step 8) times the voltage labeled on the top of the AlA3Al
V dc
EXAMPLE:
If the calibration factor calculated in step 8 is 0.00099, and AlASAl is labeled for
0.00099 x
Discharge Procedure for High Voltage and
CRT
Warning
Warning
2.
16.With the
17.Readjust AlA6R103 HV ADJ for a DVM indication equal to the voltage determined in step 15.
18.If a new CRT has just been installed do the following:
a.Set the
b.Wait an additional 30 minutes for the CRT to normalize.
c.Readjust AlA6R103 HV ADJ for a DVM indication equal to the voltage determined in step 15.
19.Set the LINE switch to STANDBY. Remove the ac line cord from each instrument section.
20.Wait at least one minute for the MAINS
21.Disconnect the
22.Replace the cover on the AlA
23.The
The
This procedure should be performed by qualified personnel only. Voltages are present which, if contacted, could cause serious personal injury. Approximately
for several days if the
1. Remove the ac line cord from both instrument sections.
With the ac power cords disconnected, voltages can still be present which, if contacted, could cause serious personal injury.
2.Obtain an electrician???s screwdriver which has a thin blade at least eight inches long. The handle of the screwdriver must be made of an insulating material.
Adjustments
2.
3.Connect one end of a jumper wire (made of insulated wire and
two alligator clips) to the blade of the screwdriver. Connect the other end of the jumper wire to the metal chassis of the IF Display Section. This grounds the screwdriver.
4.Slide the screwdriver???s blade between the CRT and the sheet metal as shown in Figure
5.Remove the cover from the AlA
6.Touch the screwdriver???s tip to the top lead of each of the 11 large vertical capacitors on the AlA
7.The AlA
G r o u n d i n g
W i r e
Figure
Note A small bracket and screw secure the AlA
3. Preliminary
Display
Adjustments (SN
3001A and Below)
Reference
Note
Note
3. Preliminary Display Adjustments (SN 3001A and Below)
AlAl Keyboard
AlA
A1 A4
AlA
Adjustment 2,
Perform this adjustment only if components have been replaced on the AlA
Equipment Digitizing Oscilloscope . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . HP 54501A Pulse/Function Generator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . HP 8116A 1O:l Divider Probe, 10 MW7.5 pF (2 required) . . . . . . . . . . . HP 10432A Display Adjustment PC Board (setice
Adjustments
3. Preliminary Display Adjustments (SN 3001A and Below)
Procedure
X and Y Deflection
Amplifier Pulse
Response Adjustments
Note
1.Connect a 10:1 (10 MQ) divider probe to the oscilloscope???s channel
1input and a 10: 1 divider probe to the channel 4 input.
2.On the oscilloscope, press CRECALL) [ml to perform a soft reset.
3.On the oscilloscope, press [CHAN] more preset probe , select channel 1, and use the
4.Select channel 4, and use the
5.Press @iYGiJ
6.Connect the channel 1 probe to the oscilloscope???s rear panel PROBE COMPENSATION AC CALIBRATOR OUTPUT connector. Press [AUTO- SCALEJ Adjust the channel 1 probe for an optimum square wave display on the oscilloscope.
7.Connect the channel 4 probe to the oscilloscope???s rear panel PROBE COMPENSATION AC CALIBRATOR OUTPUT connector. Press LAUTO- SCALE_). Adjust the channel 4 probe for an optimum square wave display on the oscilloscope.
Each probe is now compensated for the oscilloscope input to which it is connected. Do not interchange probes without recompensating.
8.Connect the channel 1 1O:l divider probe to AlA4E1,and the channel 4 probe to AlA4E2, as shown in Figure
SOQ TEPMNATIC+
PULSE/FUNCTION
GENERATOR
Figure
9.Remove the cover over A3 Digital Storage Section and remove A3A2 Intensity Control Assembly. Insert the Display Adjustment PC board (HP part number
3. Preliminary Display Adjustments (SN 3001A and Below)
A1A5,
AlA4-
AlA2???
- A3A2
Figure
~000uu000000u00ur
Figure
10. Set the Pulse/Function Generator controls as follows:
MODE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .. . . . . . . . . . NORM
Waveform . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ..pulse
Frequency (FRQ) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 200 kHz
Width(WID) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ...250 ns
Amplitude (AMP) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2.00 V
Offset (OFS) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ,000 mV
Il. Connect the output of the Pulse/Function Generator to Jl (X input) on the Display Adjustment PC board in the A3A2 slot as shown in Figure
Note The Pulse/Function Generator???s output must be terminated with 50 ohms. Use a BNC tee, a 500 termination, and a BNC female to SMB female adapter. Install the 500 termination as close to the Display Adjustment PC Board as possible.
Adjustments
3.Preliminary Display Adjustments (SN 3001A and Below)
16.Three waveforms should be displayed on the oscilloscope, as shown in Figure
oscilloscope???s
r e m o t e l i s t e n
:
Figure
17.Adjust AlA4R28 HF GAIN, AlA4Cl0, and AlA4Cll for minimum overshoot and minimum rise and fall times of the composite X deflection waveform.
Note Always adjust AlA4ClO and AlA4Cll in approximately equal amounts. Do not adjust one to its minimum value and the other to its maximum value.
18.Use the oscilloscope Intnv) markers to measure the risetime, falltime, and percent overshoot of the composite X defection waveform. Rise and fall times should both be less than approximately 65 ns between the 10% and 90% points on the waveform. Overshoot should be less than 3% (approximately 0.25 divisions). See Figure
O V E R S H O O T
O V E R S H O O T
Figure
Rise and Fall Times and Overshoot Adjustment Waveform
Adjustments
3.Preliminary Display Adjustments (SN 3001A and Below)
19.Connect the oscilloscope???s channel 1 probe to AlA5El and the channel 4 probe to AlA5E2. See Figure
20.The Y Deflection Amplifier is identical to the X Deflection Amplifier. Repeat steps 12 through 18 for the Y Deflection Amplifier using R7, R27, R28, ClO, and Cl1 respectively.
Pulse Response of Control Gate Z Amplifier to BLANK Input
21.Disconnect the oscilloscope channel 4 probe from the spectrum analyzer. Connect the oscilloscope channel 1 probe to AlA2TP2, and connect the probe???s ground lead to chassis ground.
22.On the oscilloscope, press [RECALL) (CLEAR) to perform a soft reset.
2% Press (TiTiiQ CHANNEL 1 on, more preset probe , and use the
24. Set the oscilloscope controls as follows:
Press (CHAN):
amplitude scale . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10.0 V/div offset . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ...25.0000 V
Press [TlMEBASEI):
time scale . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ..50.0ns/div delay . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 125.000 ns
Press ITRIG]:
level . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5.00000 V c
Press
connect dots . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ..on Press @ii??@.
25.Connect the output of the Pulse/Function Generator to 53 (Z input) on the Display Adjustment PC Board in the A3A2 slot. Set the board???s switch to the down position.
Note The pulse/function generator???s output must be terminated with 50 ohms. Use a BNC tee, a 5062 termination, and a BNC female to SMB female adapter. Install the 500 termination as close to the Display Adjustment PC Board as possible.
26. Set the pulse/function generator???s controls as follows:
MODE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . NORM
Waveform . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ..pulse
Frequency (FRQ) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ,200 kHz
Width(WID) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ..250ns
Amplitude (AMP) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4.OOV
Offset (OFS) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2.OOV
27.Set the spectrum analyzer???s
3.Preliminary Display Adjustments (SN 3001A and Below)
28.Set the oscilloscope controls as follows:
Press @iK]:
Channel 1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . on amplitude scale . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6.25 V/div
Press @iZZQ.
29.Adjust AlA4R7 X POS and AlA5R7 Y POS to either extreme to position the CRT beam
30.Adjust the spectrum analyzer???s
18 . 0 0 V/div offset: 4 5 . 0 0 v
10.00 : I dc
1 2 5 . 0 0 0 ns3 7 5 . 0 0 0 ns 5 0 . 0 ns/dlv
1 f 50.00 v
Figure
31.Adjust AlA2R22 HF GAIN and AIABCIO for minimum overshoot on rise and minimum rise and fall times of the pulse waveform.
32.Use the oscilloscope [ml markers to measure the risetime, falltime, and percent overshoot of the pulse waveform. Rise and falltimes should be less than 50 ns and 90 ns respectively. Overshoot on the rise should be less than 5% (approximately 0.4 divisions).
33.Set the spectrum analyzer???s LINE switch to STANDBY, and center potentiometers AlA4R7 X POSN and AlA5R7 Y POSN.
34.Disconnect the oscilloscope channel 1 probe from the spectrum analyzer. Remove the Display Adjustment PC board from the A3A2 slot, and reinstall the A3A2 Intensity Control Assembly. Replace the A3 Section cover and cables.
35.Perform Adjustment Procedure 4, Final Display Adjustment (SN 3001A and Below).
Adjustments
3. Preliminary
Display
Adjustments (SN
3004A and Above)
Reference
Note
Note
AlAl Keyboard
Al A2 X, Y, Z Axis Amplifier
Adjustment Procedure 2,
Perform this adjustment only if components have been replaced on the AlA X, Y, Z Axis Amplifier Assembly. Components R117, R217, R308, C104, C109, C204, C209, and C307 are factory adjusted and normally do not require readjustment. Components affecting these adjustments are located in function blocks F, H, M, N, 0, P, R, and S of the AlA X, Y, Z Axis Amplifier Assembly schematic diagram.
Description
Caution
The X, Y, Z Axis Amplifier Assembly is adjusted to compensate the CRT drive circuits for proper horizontal and vertical characteristics. These preliminary adjustments are necessary only after replacement or repair of the AlA X, Y, Z Axis Amplifier Assembly). For routine maintenance, CRT replacement, or minor repairs, only Adjustment Procedure 4, ???Final Display Adjustments,??? needs to be performed.
Be sure not to allow a fixed spot of high intensity to remain on the spectrum analyzer CRT. A high intensity spot may permanently damage the CRT???s phosphor coating. Monitor the CRT closely during the following adjustment procedures. If a spot occurs, move it
Equipment Digitizing Oscilloscope . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . HP 54501A Pulse/Function Generator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . HP 8116A
10:1 Divider Probe, 10 MW7.5 pF, (2 required) . . . . . . . . . . . HP 10432A Display Adjustment PC Board (service
Adapters:
Adapter, BNC(f) to
Procedure
X and Y Deflection
Amplifier Pulse
Response Adjustments
Note
3.Preliminary Display Adjustments (SN 3004A and Above)
1.Connect a 1O:l (10 MQ) divider probe to the oscilloscope???s channel
1input and a 1O:l divider probe to the channel 4 input.
2.On the oscilloscope, press (RECALL) (CLEARI) to perform a soft reset.
3.On the oscilloscope, press (CHAN) more preset probe , select channel 1, and use the
4.Select channel 4, and use the
5.Press ml.
6.Connect the channel 1 probe to the oscilloscope???s rear panel PROBE COMPENSATION AC CALIBRATOR OUTPUT connector. Press CAUTO- SCALE). Adjust the channel 1 probe for an optimum square wave display on the oscilloscope.
7.Connect the channel 4 probe to the oscilloscope???s rear panel PROBE COMPENSATION AC CALIBRATOR OUTPUT connector. Press [AUTO- SCALE]. Adjust the channel 4 probe for an optimum square wave display on the oscilloscope.
Each probe is now compensated for the oscilloscope input to which it is connected. Do not interchange probes without recompensating.
8.Connect the channel 1 IO:1 divider probe to AlA2TP204, and the channel 4 probe to AlA2TP205, as shown in Figure
5On TERkllNATlON W
Figure
9.Remove the cover over A3 Digital Storage Section and remove A3A2 Intensity Control Assembly. Insert the Display Adjustment PC board (HP part number
Adjustments
3. Preliminary Display Adjustments (SN 3004A and Above)
AlA
A3A2
A3Al
Figure
Figure
10. Set the Pulse/Function Generator controls as follows:
MODE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . NORM
Waveform . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ..pulse
Frequency (FRQ) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ,200 kHz
Width(WID) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ...250 ns
Amplitude (AMP) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2.00 V
Offset (OFS) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .O.OOO mV
11.Connect the output of the Pulse/Function Generator to Jl (X input) on the Display Adjustment PC board in the A3A2 slot as shown in Figure
Note The pulse/function generator???s output must be terminated with 50 ohms. Use a BNC tee, a 5Ofl termination, and a BNC female to SMB female adapter. Install the 5052 termination as close to the Display Adjustment PC Board as possible.
3.Preliminary Display Adjustments (SN 3004A and Above)
12.Set the oscilloscope controls as follows:
13.Set the spectrum analyzer???s
14.The X+ deflection and X- deflection waveforms should be
superimposed on the oscilloscope display, as shown in Figure
hf r u n n i n g
1 1 0 . 0 V/div offset: 2 5 . 0 0 V
10.00 : I dc
4 1 0 . 0 V/div offset: 6 0 . 0 0 v 10.00 : I dc
8..
- 1 2 5 . 0 0 0 ns
1 f 25.00 V
Figure
15. Set the oscilloscope controls as follows:
Press [WFORM MATH):
f l . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ..on display . . . . . . . . . . . .. .. .. .. .. . .. .. .. . . . . . . . . . . . . . . . . . . .. .. on math . . . . . . . . . . . . . . . .. . . . . . . . . . . channel 1 - channel 4 sensitivity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ...25.0 Vldiv
16.Three waveforms should be displayed on the oscilloscope, as shown in Figure
Adjustments
3. Preliminary Display Adjustments (SN 3004A and Above)
the combined X deflection voltage applied to the CRT. Use the oscilloscope???s
hp r u n n i n g
: 1 2 0 . 0 V/div
o f f s e t : 2 5 . 0 0 V
.I
..:..
:::. .:. ::.:::::.::.:::..
.:
- 1 2 5 . 0 0 0 ns1 2 5 . 0 0 0 ns3 7 5 . 0 0 0 ns 5 0 . 0 ns/dlv
1 f 25.00 V
Figure
17.Adjust AlA2R217 HF GAIN, AlA2C204, and AlA2C209 for minimum overshoot and minimum rise and fall times of the composite X deflection waveform.
Note Always adjust AlA2C204 and AlA2C209 in approximately equal amounts. Do not adjust one to its minimum value and the other to its maximum value.
18.Use the oscilloscope Intnv] markers to measure the risetime, falltime, and percent overshoot of the composite X defection waveform. Rise and fall times should both be less than approximately 65 ns between the 10% and 90% points on the waveform. Overshoot should be less than 3% (approximately 0.25 divisions). See Figure
O V E R S H O O T
9 0 %
O V E R S H O O T
Figure
Rise and Fall Times and Overshoot Adjustment Waveform
3.56 Adjustments
3.Preliminary Display Adjustments (SN 3004A and Above)
19.Connect the oscilloscope???s channel 1 probe to AlA2TP104 and the channel 4 probe to AlA2TP105. See Figure
20.The Y Deflection Amplifier is identical to the X Deflection Amplifier. Repeat steps 12 through 18 for the Y Deflection Amplifier using R127, R120, R117, C104, and C109, respectively.
Pulse Response of 21. Disconnect the oscilloscope channel 4 probe from the spectrum
23.Press m, CHANNEL 1 on, more preset probe , and use the
24.Set the oscilloscope controls as follows:
25.Connect the output of the Pulse/Function Generator to 53 (Z input) on the Display Adjustment PC Board in the A3A2 slot. Set the board???s switch to the down position.
Note The pulse/function generator???s output must be terminated with 50 ohms. Use a BNC tee, a 500 termination, and a BNC female to SMB female adapter. Install the 500 termination as close to the Display Adjustment PC Board as possible.
26. Set the Pulse/Function Generator???s controls as follows:
MODE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . NORM
Waveform . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ..pulse
Frequency (FRQ) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 200 kHz
Width (WID) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 250 ns
Amplitude (AMP) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4.OOV
Offset (OFS) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2.OOV
27.Disconnect the black connector with three wires (8, 98, and 96) from AlA2J5, and set AlA2R319 INT GAIN fully clockwise.
28.Set the spectrum analyzer???s
Adjustments 3.57